Skip to content

Minimed Go App (OUS iOS)

by Medtronic MINIMED

Ref MMT-8601Model MMT-8600

Active European Union. Also listed in GUDID (US).

CE MDR

Valid until

Not listed

EUDAMED holds no validity date

FDA recalls

2 ongoing

This device is named in the recall

Documents

None linked in EUDAMED

A group certificate may exist

Trust

Unverified

Registry data, not yet checked by deviCERTS. Source EUDAMED, GUDID, synced 23 Oct 2026. What this means

Regulatory status

as of registry sync 23 Oct 2026

RegionAuthorizationNumberDateStatus
European UnionCE/MDR EUDAMED Number0763000B000327286 Basic UDI-DI 00199150038961UDI-DIDate6 Jul 2026RegisteredStatusActive
United StatesGUDID listing GUDID Number00199150038961 UDI-DI Date14 Jun 2026ListedStatusListed GUDID: in commercial distribution

Certificates

0 certificates · EUDAMED

EUDAMED lists no certificate for this device's Basic UDI-DI.

A certificate issued for a device group may exist without being linked here.

Recalls

2 recalls · 2 ongoing · FDA

    • Ongoing Z-0265-2027 FDA Not yet classified initiated 2 Sep 2026

    A software defect in which existing users who had previously configured and enabled the long-acting dose reminder feature in version 1.2.0 or earlier experienced failure of the long-acting dose reminder after updating their app to version 1.2.1. The long-acting dose reminder is an optional feature that, when enabled, triggers a notification at the scheduled time to remind users to take their long-acting insulin dose. The issue was caused by a type cast exception during database migration that occurs upon app update to version 1.2.1 causing a silent failure that prevented the creation of new alerts. Failure to receive the long-acting dose reminder may result in missed or delayed insulin doses, potentially leading to hyperglycemia. Conversely, if users are uncertain whether they took their dose due to the missing reminder, inadvertent duplicate dosing may occur, potentially resulting in hypoglycemia.

    • Ongoing Z-2778-2026 FDA Class II initiated 16 Jun 2026

    For MiniMed Go app version 1.0.0 and 1.1.0, when the Set day and night is turned OFF and missed dose or Correct high glucose alerts are enabled, the associated alerts will not be triggered during the nighttime. This could result in missed rapid-acting insulin doses without the user s knowledge, and may lead to delayed therapy and an increased risk of hyperglycemia.

Coverage

2 jurisdictions

United States of AmericaSwedenFinlandLatviaEstoniaDenmarkLithuaniaIrelandNetherlandsGermanyPolandFranceBelgiumLuxembourgCzech RepublicSlovakiaAustriaHungaryRomaniaSloveniaBulgariaPortugalSpainItalyCroatiaGreeceMaltaCyprus
  • Active
  • Expiring
  • Withdrawn
  • Listed
  • Not covered
  • European Union CE/MDR Active
  • United States GUDID listing Listed

Specification

EUDAMED · GUDID

Trade Name
MiniMed Go
Generic Name
Percutaneous interstitial fluid glucose monitoring system, electrochemical
Device Name
Minimed Go App (OUS iOS)
Model / Reference
MMT-8600, MMT-8601
Primary DI / UDI-DI
00199150038961 EUDAMED
00199150038985 GUDID
Basic UDI-DI
0763000B000327286
510(k) Number
FDA Product Code
NDC
EMDN Code
W020106019099 Various diabetes monitoring instruments - other
GMDN Term
Percutaneous interstitial fluid glucose monitoring system, electrochemical
Device Class
Class IIa EUDAMED
Class II GUDID
Regulation Number
868.1890
Special Device Type
Software
EU market of registration
Netherlands
GUDID distribution status
In Commercial Distribution
MRI Safety
MR Unsafe

Truth per registry

2 sources · 2 fields differ

FieldEUDAMEDGUDID
Primary DI / UDI-DIEUDAMED
00199150038961
GUDID
00199150038985
Device ClassEUDAMED
Class IIa
GUDID
Class II

Manufacturer details

EUDAMED SRN
US-MF-000023100
Authorised representative
Medtronic B.V. NL-AR-000006050

Other devices from Medtronic Minimed, Inc.

See all devices from Medtronic Minimed, Inc. →

Properties

  • Over-the-Counter
  • Lot / Batch Number

Other properties (22)

  • Sterile: No
  • Single Use: No
  • Implantable: No
  • Reusable: No
  • Contains Latex: No
  • Reprocessed: No
  • CMR Substance: No
  • Human-derived: No
  • Human Tissues: No
  • Animal Tissues: No
  • Endocrine Disruptor: No
  • Administering Medicine: No
  • Measuring Function: No
  • Incorporates Medicinal Product: No
  • Prescription Use: No
  • Convenience Kit: No
  • Combination Product: No
  • HCT/P: No
  • Premarket Exempt: No
  • Serial Number: No
  • Expiration Date: No
  • Manufacturing Date: No