Skip to content

Mobicath Bi-Directional Guiding Sheath Model 1000182-Xxx

by Great Batch Medical

510(k) K101784

Active United States

510(k)

Valid until

No expiry

510(k) clearance is not time-limited

FDA recalls

None ongoing

3 closed

Documents

Trust

Unverified

Registry data, not yet checked by deviCERTS. Source FDA 510(k), synced 22 Oct 2026. What this means

Regulatory status

as of registry sync 22 Oct 2026

RegionAuthorizationNumberDateStatus
United StatesFDA 510(k)SESE Substantially Equivalent · Traditional NumberK101784 Date26 Nov 2010ClearedStatusActive

Recalls

3 recalls · FDA

    • Terminated Z-2914-2016 FDA Class II initiated 19 Sep 2016 on shared 510(k) K101784

    FDA inspection found that customers who had received devices with promotional material with outside use indications were not notified when labeling was revised to remove these indications.

    • Terminated Z-0611-2013 FDA Class II initiated 12 Nov 2012

    Greatbatch Medical observed some anomalies of the inner lumen of the MobiCath Bi-Directional Guiding Sheaths upon implementation of a new visual inspection tool. The anomalies may include loose, string-like liner material, liner abrasion (fraying), and/or attached string-like liner material (from scratches or damage on the inner wall of sheath lumen).

    • Terminated Z-0132-2013 FDA Class II initiated 24 Sep 2012

    We are notifying you of a potential interaction between therapeutic catheters and MobiCath Bi-Directional Guiding Sheaths. We became aware of these interactions while investigating recent complaints regarding catheters "self-advancing" while being used in conjunction with the MobiCath Sheaths.

Recalls on a shared submission name another device cleared under the same FDA submission as this one.

Coverage

1 jurisdiction

United States of America
  • Active
  • Expiring
  • Withdrawn
  • Listed
  • Not covered

United States — FDA 510(k) Active

Specification

510(k)

510(k) Number
FDA Product Code
DYB
Device Class
Class II
Regulation
21 CFR Part 870.1340
Advisory Committee
Cardiovascular
FDA Decision
SESE Substantially Equivalent

Other devices from Great Batch Medical

See all devices from Great Batch Medical →