Mobicath Bi-Directional Guiding Sheath Model 1000182-Xxx
510(k) K101784
Active
Valid until
No expiry
FDA recalls
None ongoingDocuments
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK101784 | Date26 Nov 2010 | StatusActive |
Recalls
-
- Terminated Z-2914-2016
FDA inspection found that customers who had received devices with promotional material with outside use indications were not notified when labeling was revised to remove these indications.
-
- Terminated Z-0611-2013
Greatbatch Medical observed some anomalies of the inner lumen of the MobiCath Bi-Directional Guiding Sheaths upon implementation of a new visual inspection tool. The anomalies may include loose, string-like liner material, liner abrasion (fraying), and/or attached string-like liner material (from scratches or damage on the inner wall of sheath lumen).
-
- Terminated Z-0132-2013
We are notifying you of a potential interaction between therapeutic catheters and MobiCath Bi-Directional Guiding Sheaths. We became aware of these interactions while investigating recent complaints regarding catheters "self-advancing" while being used in conjunction with the MobiCath Sheaths.
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA Product Code
- DYB
- Device Class
- Class II
- Regulation
- 21 CFR Part 870.1340
- Advisory Committee
- Cardiovascular
- FDA Decision
- SESE Substantially Equivalent