Mobilediagnost Wdr
510(k) K111725
Mobilediagnost Wdr is a medical device (FDA Class II; FDA product code IZL, System, X-Ray, Mobile) made by Sedecal., Sa., cleared under FDA 510(k) K111725 on 19 Jul 2011, active as of 22 Oct 2026. 8 FDA recalls are linked to it.
Active
Valid until
No expiry
FDA recalls
2 ongoingDocuments
Trust
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK111725 | Date19 Jul 2011 | StatusActive |
Official records
- FDA 510(k) database K111725
Recalls
-
- Ongoing Z-0243-2026
Sedecal sent an email to its consignees to remind the users that the equipment is not water-resistant and must be cleaned strictly according to the instructions in the manual, clearly outlining the possible consequences of improper cleaning.
-
- Ongoing Z-1083-2024
While cleaning or disinfecting the system, if the process is not followed as described in the manual and excessive liquid is used or sprayed directly on the system, it could cause a short circuit of the fine positioning control button. This would result in the system moving on its own in a very slow speed forward circular motion.
-
- Terminated Z-1084-2016
This is related to previous 806 Notifications about uncommanded movements of a motorized mobile x-ray system. Reason for Recall: Along with the previously noted encoder and gauges problems and corresponding fixes, we have observed that static electricity can also elicit the unwanted behavior.
-
- Terminated Z-1691-2016
Due to a software defect, the system may sporadically apply the default x ray exposure parameters for an adult ( patient type : Normal ) even though the patient type " Newborn" was selected and is displayed in the generator control area of the Eleva User Interface.
-
- Terminated Z-1040-2015
Reports of unexpected movement.
-
- Terminated Z-0508-2015
In special cases, during the start-up of the current model of MobileDiagnost wDR(WmDR 1.1), an unintended exposure could occur.
-
- Terminated Z-1964-2013
The Instructions for Use (IFU) for the MobileDiagnost wDR fails to comply with a Federal standard. The IFU does not provide tolerance levels for several generator values as required by 1020.30(h)(3).
-
- Terminated Z-1706-2017
When the operator moves an image to another view then the electronic marker might be placed at the wrong anatomical position in the resulting image.
Timeline
- Cleared FDA 510(k) K111725
- Recall initiated FDA recalls Z-1706-2017
- Recall initiated FDA recalls Z-1964-2013
- Recall initiated FDA recalls Z-0508-2015
- Recall initiated FDA recalls Z-1040-2015
- Recall initiated FDA recalls Z-1691-2016
- Recall initiated FDA recalls Z-1084-2016
- Recall initiated FDA recalls Z-1083-2024
- Recall initiated FDA recalls Z-0243-2026
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA product code
- IZL
- Device class
- Class II
- Regulation
- 21 CFR Part 892.1720
- Advisory committee
- Radiology
- FDA decision
- SESE Substantially Equivalent
Similar devices
Other devices filed under FDA product code IZL (System, X-Ray, Mobile).
- PRORAD ATLAS ULTRAPORTABLE; PRORAD ATLAS ULTRAPORTABLE PLUS by Prognosys Medical Systems Private Limited
- AMX4-50 AND AMX4-31 MOBILE DIGITAL DIAGNOSTIC RADIOGRAPHIC SYSTEMS by Virtual Imaging, Inc.
- Amadeo M-DR Mini, Amadeo M-AX Mini by OEHM UND REHBEIN GMBH
- MINE ALNU (Model: Mine 2.1) by Hdt Co., Ltd.
- Micro C Medical Imaging System, M01 by Oxos Medical
- RAYMO Mobile X-ray System (Model: RAYMO) by DRGEM Corporation
- EzRay M18 (Model: VMX-P400) by VATECH CO., LTD.
- BETA I & BETA II by Porta Ray, Inc.
- XLITE by GEMSS HEALTHCARE CO., LTD.
- CP 100/30 by Micro-Focus Imaging., Inc.