Model 780
by Hudson Respiratory Care, Inc.
510(k) K031383
Active
Valid until
No expiry
FDA recalls
1 ongoingDocuments
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK031383 | Date7 Jul 2003 | StatusActive |
Recalls
-
- Ongoing Z-2991-2026
Neonatal and infant heated wire breathing circuits may experience sporadic failure. The heated wire connector (pigtail) at the humidifier may exhibit discoloration, melting, smoke, burning odor, or other signs of localized thermal damage. Device failure may predispose infant to bronchospasm, inspissated secretions, and airway obstruction, as well as inadvertent positive end-expiratory pressure (PEEP), circuit obstruction, or accidental lavage of pooled fluid into the airway, elevated oxygen consumption, and worsening respiratory distress.
-
- Terminated Z-2107-2014
The temperature probe does not properly connect to the temperature port.
-
- Terminated Z-0419-2016
Product is being recalled due to complaints that the inspiratory and expiratory tubes are transposed.
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA Product Code
- BTT
- Device Class
- Class II
- Regulation
- 21 CFR Part 868.5450
- Advisory Committee
- Anesthesiology
- FDA Decision
- SESE Substantially Equivalent