# Modification to ABL800 Flex

> Active: United States. Medical device record on deviCERTS, compiled from public registries.

- Canonical page: <https://devicerts.com/devices/modification-to-abl800-flex-cig>
- Manufacturer: [Radiometer Medical ApS](https://devicerts.com/manufacturers/radiometer-medical-aps) (FDA registration: Denmark)
- Name in the registry: MODIFICATION TO ABL800 FLEX
- Device class: Class II (US FDA)
- FDA product code: CIG

## Registrations and authorizations

- **510(k)** · FDA 510(k) · United States · K050869 · decision 2005-08-11 · decision SESE · status: Active

## Recalls

- Z-0115-2007 ·  · terminated · initiated 2006-07-06 · matched through shared submission K050869, may concern other models: Incorrect FI02 values: programming issue can cause incorrect values to be transmitted to the LIS when : 1)- The FIO2 result is edited in the manual sample processing mode--2)- An existing result is opened and FI02 is then edited and sent.....In both cases RADIANCE will transmit the original FI02 value, not the value that was edited.

## Identifiers

- 510(k) Number: K050869 [FDA 510(k)]
- FDA product code: CIG [FDA 510(k)]

## Classification

- Device class: Class II [FDA 510(k)]
- Regulation: 21 CFR Part 862.1110 [FDA 510(k)]
- Advisory committee: Clinical Chemistry [FDA 510(k)]
- FDA decision: SESE [FDA 510(k)]

## Source and citation

- Cite as: Modification to ABL800 Flex, Radiometer Medical ApS. deviCERTS medical device catalogue. https://devicerts.com/devices/modification-to-abl800-flex-cig (data as of 2026-11-17).
- Data sources: [FDA 510(k)](https://open.fda.gov/apis/device/510k/), [FDA recalls](https://open.fda.gov/apis/device/recall/)
- Registry records last synced: 2026-10-22
- Statuses computed as of: 2026-11-17
- JSON: <https://devicerts.com/devices/modification-to-abl800-flex-cig.json>
- Licence: [CC BY 4.0](https://creativecommons.org/licenses/by/4.0/). The deviCERTS layout and text are licensed under CC BY 4.0. Registry data remain under their source terms: U.S. FDA data (openFDA, AccessGUDID) are U.S. public domain; EUDAMED data are subject to the European Commission reuse terms.
- Attribution: DevicErts, https://devicerts.com/devices/modification-to-abl800-flex-cig
- Compiled by deviCERTS (devicerts.com) from public medical device registries (EUDAMED, European Commission; U.S. FDA via openFDA; AccessGUDID, U.S. National Library of Medicine). deviCERTS is not an official register and gives no regulatory advice: the record in the registry itself prevails.