Modification to Avea Ventilator
by Viasys Respiratory Care, Inc.
510(k) K073069
Active
Valid until
No expiry
FDA recalls
None ongoingDocuments
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK073069 | Date25 Jan 2008 | StatusActive |
Recalls
-
- Terminated Z-1894-2016
- Terminated Z-1888-2016
- Terminated Z-1892-2016
- Terminated Z-1897-2016
- Terminated Z-1895-2016
- Terminated Z-1896-2016
- Terminated Z-1898-2016
- Terminated Z-1899-2016
- Terminated Z-1889-2016
- Terminated Z-1890-2016
- Terminated Z-1893-2016
- Terminated Z-1891-2016
CareFusion has identified a potential risk associated with AVEA Ventilator caused by an incorrect F1 fuse on the AVEA Ventilator Transducer Communication Alarm (TCA) Board which may fail/blow, causing loss of power to the User Interface Module (UIM).
-
- Terminated Z-2273-2013
CareFusion has identified a potential risk associated with AVEA ventilators when used at higher altitudes. A proactive complaint review identified an error in the barometric pressure compensation. CareFusion is voluntarily initiating a field correction of the affected devices to correct the error.
-
- Terminated Z-0510-2012
The recall was initiated because Carefusion has identified a potential risk associated with certain AVEA ventilators and affected replacement parts manufactured between March 1, 2009 and June 30, 2011. Carefusion is voluntarily initiating a field correction of the affected devices to preclude the possibility of this risk.
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA Product Code
- CBK
- Device Class
- Class II
- Regulation
- 21 CFR Part 868.5895
- Advisory Committee
- Anesthesiology
- FDA Decision
- SESE Substantially Equivalent