Recalls
-
- Terminated Z-2030-2010
Medtronic received reports involving the V100D U-Clip where during handling, the release mechanism has become separated from the needle assembly. These instances of separation have occurred when pulling the clip through prosthetic material, or when grasping and releasing the clip using robotic instrumentation. While this issue does not impact performance of the clip once implanted, separation of
-
- Terminated Z-1609-2008
Medtronic has identified one lot #0610315 of U-Clip Removal Tools, Catalog UREM1860FW, that was incorrectly assembled. The nose cone attachment, on the body of the removal tool, may not contain sufficient adhesive, which can allow the nose cone to separate from the body of the device.
-
- Terminated Z-0968-2007
Six lots of U-Clip Removal Tools, catalog REM1860FW were incorrectly assembled. The nose cone attachment to the body of the removal tool does not contain sufficient adhesive which can cause the nose to separate from the body of the device.
-
- Terminated Z-0763-06
Customers stated that the U-Clip deployment collar was either difficult to rotate or 'locks up' during usage, which prevents the deployment of the U-Clips.
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA Product Code
- FZP
- Device Class
- Class II
- Regulation
- 21 CFR Part 878.4300
- Advisory Committee
- General, Plastic Surgery
- FDA Decision
- SESE Substantially Equivalent