Modification to Focus
510(k) K022630
Active
Valid until
No expiry
FDA recalls
None ongoingDocuments
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK022630 | Date9 Oct 2002 | StatusActive |
Recalls
-
- Terminated Z-1091-2014
Instrumentarium Dental has delivered a number of FOCUS 3.0 Intraoral X-ray units in November 2013 with a potential flaw in the mounting hardware. A number of wall mount units were delivered with FOCUS X-ray machines that had not been properly welded at our supplier. As a result of the incomplete welding operation, the wall mount unit on the FOCUS X-ray unit may not properly support the unit as expected, which could cause the unit to drift. Depending on the extent of the shortened or missing welds, there is also a possibility that the wall mount unit will fail to support the weight of the unit, resulting in the FOCUS X-ray unit falling unexpectedly from its mounted position.
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA Product Code
- EHD
- Device Class
- Class II
- Regulation
- 21 CFR Part 872.1800
- Advisory Committee
- Dental
- FDA Decision
- SESE Substantially Equivalent