Modification to Neurosensor Cerebral Blood Flow and Intracranial Pressure Monitoring System
by Integra Lifesciences Corporation
510(k) K071505
Active
510(k)
Valid until
No expiry
FDA recalls
None ongoingDocuments
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK071505 | Date6 Aug 2007 | StatusActive |
Recalls
-
- Terminated Z-1191-2008
Incorrect pressure reading: Due to a manufacturing error, the ICP reading could be inaccurate to the extent that they exceed label claims. Specifically, the Insertion Error label claim of +/- 3mmHg could be exceeded by up to -3mmHg, that could result in a maximum Insertion Error of +/- 6mmHg.
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA Product Code
- GWM
- Device Class
- Class II
- Regulation
- 21 CFR Part 882.1620
- Advisory Committee
- Neurology
- FDA Decision
- SESE Substantially Equivalent