Skip to content

Modification to Neurosensor Cerebral Blood Flow and Intracranial Pressure Monitoring System

by Integra Lifesciences Corporation

510(k) K071505

Active United States

510(k)

Valid until

No expiry

510(k) clearance is not time-limited

FDA recalls

None ongoing

1 closed

Documents

Trust

Unverified

Registry data, not yet checked by deviCERTS. Source FDA 510(k), synced 22 Oct 2026. What this means

Regulatory status

as of registry sync 22 Oct 2026

RegionAuthorizationNumberDateStatus
United StatesFDA 510(k)SESE Substantially Equivalent · Traditional NumberK071505 Date6 Aug 2007ClearedStatusActive

Recalls

1 recall · FDA

    • Terminated Z-1191-2008 FDA Unclassified (pre-2012) initiated 28 Jan 2008

    Incorrect pressure reading: Due to a manufacturing error, the ICP reading could be inaccurate to the extent that they exceed label claims. Specifically, the Insertion Error label claim of +/- 3mmHg could be exceeded by up to -3mmHg, that could result in a maximum Insertion Error of +/- 6mmHg.

Coverage

1 jurisdiction

United States of America
  • Active
  • Expiring
  • Withdrawn
  • Listed
  • Not covered

United States — FDA 510(k) Active

Specification

510(k)

510(k) Number
FDA Product Code
GWM
Device Class
Class II
Regulation
21 CFR Part 882.1620
Advisory Committee
Neurology
FDA Decision
SESE Substantially Equivalent

Other devices from Integra Lifesciences Corporation

See all devices from Integra Lifesciences Corporation →