Skip to content

Modification to Outback Catheter

by Lumend, Inc. 4 variants

510(k) K040771

Active United States

510(k)

Valid until

No expiry

510(k) clearance is not time-limited

FDA recalls

None ongoing

1 closed

Documents

Trust

Unverified

Registry data, not yet checked by deviCERTS. Source FDA 510(k), synced 22 Oct 2026. What this means

Regulatory status

as of registry sync 22 Oct 2026

RegionAuthorizationNumberDateStatus
United StatesFDA 510(k)SESE Substantially Equivalent · Special NumberK040771 Date23 Apr 2004ClearedStatusActive

Recalls

1 recall · FDA

    • Terminated Z-0279-2009 FDA Unclassified (pre-2012) initiated 20 Aug 2008 on shared 510(k) K040771

    The Cordis LuMend Outback LTD Re-Entry Catheter - separation rate of the cannula to the deployment slide of the handle assembly is higher than anticipated. The cannula is unable to be retracted into the device after deployment due to a separation of the inner key from the cannula.

Recalls on a shared submission name another device cleared under the same FDA submission as this one.

Coverage

1 jurisdiction

United States of America
  • Active
  • Expiring
  • Withdrawn
  • Listed
  • Not covered

United States — FDA 510(k) Active

Specification

510(k)

510(k) Number
FDA Product Code
DQY
Device Class
Class II
Regulation
21 CFR Part 870.1250
Advisory Committee
Cardiovascular
FDA Decision
SESE Substantially Equivalent

Other devices from Lumend, Inc.