  "format": "devicerts.device.v1",
  "cite_as": "Modified Distal Radius Plating System, Biomet. deviCERTS medical device catalogue. https://devicerts.com/devices/modified-distal-radius-plating-system-hrs (data as of 2026-11-24).",
  "source": {
    "canonical_url": "https://devicerts.com/devices/modified-distal-radius-plating-system-hrs",
    "json_url": "https://devicerts.com/devices/modified-distal-radius-plating-system-hrs.json",
    "markdown_url": "https://devicerts.com/devices/modified-distal-radius-plating-system-hrs.md",
    "publisher": "deviCERTS (devicerts.com)",
    "data_sources": [
      {
        "registry": "FDA 510(k)",
        "url": "https://open.fda.gov/apis/device/510k/"
      },
      {
        "registry": "FDA recalls",
        "url": "https://open.fda.gov/apis/device/recall/"
      }
    ],
    "registry_records_last_synced": "2026-10-22",
    "status_as_of": "2026-11-24",
    "license": "CC BY 4.0",
    "license_url": "https://creativecommons.org/licenses/by/4.0/",
    "license_scope": "The deviCERTS layout and text are licensed under CC BY 4.0. Registry data remain under their source terms: U.S. FDA data (openFDA, AccessGUDID) are U.S. public domain; EUDAMED data are subject to the European Commission reuse terms.",
    "attribution": "DevicErts, https://devicerts.com/devices/modified-distal-radius-plating-system-hrs",
    "disclaimer": "Compiled by deviCERTS (devicerts.com) from public medical device registries (EUDAMED, European Commission; U.S. FDA via openFDA; AccessGUDID, U.S. National Library of Medicine). deviCERTS is not an official register and gives no regulatory advice: the record in the registry itself prevails."
  },
  "device": {
    "name": "Modified Distal Radius Plating System",
    "registry_name": "MODIFIED DISTAL RADIUS PLATING SYSTEM",
    "url": "https://devicerts.com/devices/modified-distal-radius-plating-system-hrs",
    "manufacturer": {
      "name": "Biomet",
      "url": "https://devicerts.com/manufacturers/biomet-inc",
      "country": "United States",
      "country_basis": "FDA 510(k) applicant",
      "fda_registration_country": null
    },
    "status": {
      "state": "active",
      "label": "Active",
      "detail": "United States"
    },
    "classification": [
      {
        "class": "Class II",
        "system": "US FDA"
      }
    ],
    "identifiers": [
      {
        "label": "510(k) Number",
        "values": [
          {
            "value": "K133939",
            "registries": [
              "FDA 510(k)"
            ]
          }
        ]
      }
    ],
    "product_codes": [
      "HRS"
    ],
    "emdn": [],
    "gmdn": [],
    "description": null,
    "noindex": false
  },
  "authorizations": [
    {
      "registry": "510(k)",
      "region": "United States",
      "type": "FDA 510(k)",
      "number": "K133939",
      "number_label": null,
      "device_udi_di": null,
      "date": "2014-04-01",
      "date_kind": "decision",
      "decision_code": "SESE",
      "expires": null,
      "withdrawn": null,
      "status": "Active",
      "status_note": null,
      "derived_from_gudid": false
    }
  ],
  "no_longer_listed_by": [],
  "certificates": [],
  "recalls": [
    {
      "number": "Z-0810-2022",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2022-02-14",
      "reason": "Packaging of various implants may not have sufficient adhesion on all sides, leading to loss of sterile barrier integrity. Use of product poses risk of infection, leading to surgical intervention, and/or extension of surgery to find another part.",
      "applies": "shared submission",
      "via": "K133939"
    },
    {
      "number": "Z-1082-2016",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2016-01-29",
      "reason": "Components of the DVR ePAKs can oxidize during shipment and storage prior to use.  A delay in surgery, less than 30 minutes, may occur if the surgical staff proceeds to re-sterilize the product. \nIf oxidative material would flake off a component contained within the ePAK and enter the surgical field/wound an inflammatory reaction may occur. The inflammatory reaction may result in a foreign body reaction and/or metal hypersensitivity.  Oxidized components have failed cytotoxicity testing.",
      "applies": "shared submission",
      "via": "K133939"
    }
  ],
  "eu_market_availability": null,
  "properties": [
    {
      "group": "Identifiers",
      "fields": [
        {
          "label": "510(k) Number",
          "values": [
            {
              "value": "K133939",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        },
        {
          "label": "FDA product code",
          "values": [
            {
              "value": "HRS",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        }
      ]
    },
    {
      "group": "Classification",
      "fields": [
        {
          "label": "Device class",
          "values": [
            {
              "value": "Class II",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        },
        {
          "label": "Regulation",
          "values": [
            {
              "value": "21 CFR Part 888.3030",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        },
        {
          "label": "Advisory committee",
          "values": [
            {
              "value": "Orthopedic",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        },
        {
          "label": "FDA decision",
          "values": [
            {
              "value": "SESE",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        }
      ]
    }
  ]
}