# Modified Distal Radius Plating System

> Active: United States. Medical device record on deviCERTS, compiled from public registries.

- Canonical page: <https://devicerts.com/devices/modified-distal-radius-plating-system-hrs>
- Manufacturer: [Biomet](https://devicerts.com/manufacturers/biomet-inc) (United States, FDA 510(k) applicant)
- Name in the registry: MODIFIED DISTAL RADIUS PLATING SYSTEM
- Device class: Class II (US FDA)
- FDA product code: HRS

## Registrations and authorizations

- **510(k)** · FDA 510(k) · United States · K133939 · decision 2014-04-01 · decision SESE · status: Active

## Recalls

- Z-0810-2022 · Class II · terminated · initiated 2022-02-14 · matched through shared submission K133939, may concern other models: Packaging of various implants may not have sufficient adhesion on all sides, leading to loss of sterile barrier integrity. Use of product poses risk of infection, leading to surgical intervention, and/or extension of surgery to find another part.
- Z-1082-2016 · Class II · terminated · initiated 2016-01-29 · matched through shared submission K133939, may concern other models: Components of the DVR ePAKs can oxidize during shipment and storage prior to use. A delay in surgery, less than 30 minutes, may occur if the surgical staff proceeds to re-sterilize the product. If oxidative material would flake off a component contained within the ePAK and enter the surgical field/wound an inflammatory reaction may occur. The inflammatory reaction may result in a foreign body reaction and/or metal hypersensitivity. Oxidized components have failed cytotoxicity testing.

## Identifiers

- 510(k) Number: K133939 [FDA 510(k)]
- FDA product code: HRS [FDA 510(k)]

## Classification

- Device class: Class II [FDA 510(k)]
- Regulation: 21 CFR Part 888.3030 [FDA 510(k)]
- Advisory committee: Orthopedic [FDA 510(k)]
- FDA decision: SESE [FDA 510(k)]

## Source and citation

- Cite as: Modified Distal Radius Plating System, Biomet. deviCERTS medical device catalogue. https://devicerts.com/devices/modified-distal-radius-plating-system-hrs (data as of 2026-11-24).
- Data sources: [FDA 510(k)](https://open.fda.gov/apis/device/510k/), [FDA recalls](https://open.fda.gov/apis/device/recall/)
- Registry records last synced: 2026-10-22
- Statuses computed as of: 2026-11-24
- JSON: <https://devicerts.com/devices/modified-distal-radius-plating-system-hrs.json>
- Licence: [CC BY 4.0](https://creativecommons.org/licenses/by/4.0/). The deviCERTS layout and text are licensed under CC BY 4.0. Registry data remain under their source terms: U.S. FDA data (openFDA, AccessGUDID) are U.S. public domain; EUDAMED data are subject to the European Commission reuse terms.