Modified Trevo Provue Retriever
by Concentric Medical, Inc. 2 variants
510(k) K133464
Active
510(k)
Valid until
No expiry
FDA recalls
1 ongoingDocuments
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK133464 | Date17 Jan 2014 | StatusActive |
Recalls
-
- Ongoing Z-1903-2024
Neurovascular devices did not meet the minimum sample size required for endotoxin testing during manufacturing processes and are out of compliance with the AAMI ST72 standard, so these devices may exceed the acceptable limits of endotoxin levels; high levels of endotoxins may increase the risk of fever, shock and aseptic meningitis.
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA Product Code
- NRY
- Device Class
- Class II
- Regulation
- 21 CFR Part 870.1250
- Advisory Committee
- Cardiovascular
- FDA Decision
- SESE Substantially Equivalent