Skip to content

Modular Hybrid Glenoid

by Biomet Manufacturing Corp

510(k) K060694

Active United States

510(k)

Valid until

No expiry

510(k) clearance is not time-limited

FDA recalls

None ongoing

4 closed

Documents

Trust

Unverified

Registry data, not yet checked by deviCERTS. Source FDA 510(k), synced 22 Oct 2026. What this means

Regulatory status

as of registry sync 22 Oct 2026

RegionAuthorizationNumberDateStatus
United StatesFDA 510(k)SESE Substantially Equivalent · Traditional NumberK060694 Date11 Dec 2006ClearedStatusActive

Recalls

4 recalls · FDA

    • Terminated Z-2216-2017 FDA Class II initiated 26 Apr 2017

    Zimmer Biomet is conducting a voluntary recall for a single lot of the Modular Hybrid Glenoid Base Large for the Comprehensive Shoulder System after it was determined that two pieces from the lot were moved to finished goods and shipped despite having an open Non-Conformance Report (NCR). The NCR indicated that one piece from the lot had scratches and the other non-confirming threads.

    • Terminated Z-1859-2017 FDA Class II initiated 17 Feb 2017

    Endotoxin levels higher than process maximum limits were discovered in the identified polyethylene components.

    • Terminated Z-1658-2008 FDA Unclassified (pre-2012) initiated 13 Mar 2008 on shared 510(k) K060694

    The glenoid implant may not seat fully on the face of the natural glenoid due to inadequate measurement tolerances (tolerance stack-up) in the design of the cutter. Use of the instrument may lead to incomplete seating of the glenoid implant, elevating the theoretical risk of glenoid loosening, which may lead to early revision surgery. The problem may contribute to early device failure and patien

    • Terminated Z-0660-2013 FDA Class II initiated 6 Dec 12

    Biomet is recalling Part Number PT-113950 PT Hybrid Glenoid Post, following an investigation which identified that the male thread of the post may be oversized. This oversized condition can vary in degree and may lead to the following three events: 1) If the PT Hybrid Glenoid Post is not fully seated into the Hybrid Base and the implant construct is implanted, then a gap of 1-3 mm will be prese

Recalls on a shared submission name another device cleared under the same FDA submission as this one.

Coverage

1 jurisdiction

United States of America
  • Active
  • Expiring
  • Withdrawn
  • Listed
  • Not covered

United States — FDA 510(k) Active

Specification

510(k)

510(k) Number
FDA Product Code
KWS
Device Class
Class II
Regulation
21 CFR Part 888.3660
Advisory Committee
Orthopedic
FDA Decision
SESE Substantially Equivalent

Other devices from Biomet Manufacturing Corp