  "format": "devicerts.device.v1",
  "cite_as": "Monaco, Impac Medical Systems, Inc. deviCERTS medical device catalogue. https://devicerts.com/devices/monaco-muj-k151233 (data as of 2026-11-16).",
  "source": {
    "canonical_url": "https://devicerts.com/devices/monaco-muj-k151233",
    "json_url": "https://devicerts.com/devices/monaco-muj-k151233.json",
    "markdown_url": "https://devicerts.com/devices/monaco-muj-k151233.md",
    "publisher": "deviCERTS (devicerts.com)",
    "data_sources": [
      {
        "registry": "FDA 510(k)",
        "url": "https://open.fda.gov/apis/device/510k/"
      },
      {
        "registry": "FDA recalls",
        "url": "https://open.fda.gov/apis/device/recall/"
      }
    ],
    "registry_records_last_synced": "2026-10-22",
    "status_as_of": "2026-11-16",
    "license": "CC BY 4.0",
    "license_url": "https://creativecommons.org/licenses/by/4.0/",
    "license_scope": "The deviCERTS layout and text are licensed under CC BY 4.0. Registry data remain under their source terms: U.S. FDA data (openFDA, AccessGUDID) are U.S. public domain; EUDAMED data are subject to the European Commission reuse terms.",
    "attribution": "DevicErts, https://devicerts.com/devices/monaco-muj-k151233",
    "disclaimer": "Compiled by deviCERTS (devicerts.com) from public medical device registries (EUDAMED, European Commission; U.S. FDA via openFDA; AccessGUDID, U.S. National Library of Medicine). deviCERTS is not an official register and gives no regulatory advice: the record in the registry itself prevails."
  },
  "device": {
    "name": "Monaco",
    "registry_name": "Monaco",
    "url": "https://devicerts.com/devices/monaco-muj-k151233",
    "manufacturer": {
      "name": "Impac Medical Systems, Inc.",
      "url": "https://devicerts.com/manufacturers/impac-medical-systems-inc",
      "country": "United States",
      "country_basis": "FDA 510(k) applicant",
      "fda_registration_country": null
    },
    "status": {
      "state": "active",
      "label": "Active",
      "detail": "United States"
    },
    "classification": [
      {
        "class": "Class II",
        "system": "US FDA"
      }
    ],
    "identifiers": [
      {
        "label": "510(k) Number",
        "values": [
          {
            "value": "K151233",
            "registries": [
              "FDA 510(k)"
            ]
          }
        ]
      }
    ],
    "product_codes": [
      "MUJ"
    ],
    "emdn": [],
    "gmdn": [],
    "description": null,
    "noindex": false
  },
  "authorizations": [
    {
      "registry": "510(k)",
      "region": "United States",
      "type": "FDA 510(k)",
      "number": "K151233",
      "number_label": null,
      "device_udi_di": null,
      "date": "2015-07-07",
      "date_kind": "decision",
      "decision_code": "SESE",
      "expires": null,
      "withdrawn": null,
      "status": "Active",
      "status_note": null,
      "derived_from_gudid": false
    }
  ],
  "no_longer_listed_by": [],
  "certificates": [],
  "recalls": [
    {
      "number": "Z-1233-2023",
      "classification": "Class II",
      "status": "ongoing",
      "initiated": "2023-02-28",
      "reason": "Re-optimization, after adding contours without forced density outside the external structure, may result in inaccurate dose presentation.",
      "applies": "this device",
      "via": "K151233"
    },
    {
      "number": "Z-2091-2019",
      "classification": "Class II",
      "status": "ongoing",
      "initiated": "2019-07-03",
      "reason": "When creating 3D plans using either MU or Dose weighting modes, if the user changes the number of fractions, rescales the plan and then changes the wedge angle, the Monitor Units are scaled incorrectly.",
      "applies": "this device",
      "via": "K151233"
    },
    {
      "number": "Z-1044-2017",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2017-01-10",
      "reason": "Incorrect dose after editing beam number an wedge angle.",
      "applies": "this device",
      "via": "K151233"
    },
    {
      "number": "Z-1009-2017",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2016-12-21",
      "reason": "Incorrect Enhanced Dynamic Wedge (EDW) or Virtual Wedge (VW) Calculations.",
      "applies": "this device",
      "via": "K151233"
    },
    {
      "number": "Z-0660-2017",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2016-11-03",
      "reason": "Incorrect Dose when using the reset function.",
      "applies": "this device",
      "via": "K151233"
    },
    {
      "number": "Z-0076-2017",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2016-09-29",
      "reason": "When creating 3D plans using either MU or Dose weighting modes, if the user changes the Physician's Intent Rx Dose and/or the number of fractions, and then modifies the wedge angle, the MU value is scaled incorrectly.",
      "applies": "this device",
      "via": "K151233"
    },
    {
      "number": "Z-1879-2016",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2016-05-18",
      "reason": "When DICOM exporting a 3D Monaco plan and the \"Composite Field Sequencing\" (CFS) checkbox is selected, the Y jaws will snap to the port shape when they should remain where they were defined.",
      "applies": "this device",
      "via": "K151233"
    },
    {
      "number": "Z-0181-2016",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2015-10-16",
      "reason": "Unintended update of Dose and MU and Incorrect Assignment of Bolus.",
      "applies": "this device",
      "via": "K151233"
    },
    {
      "number": "Z-0112-2016",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2015-09-15",
      "reason": "Dose and MU are incorrect when CT images are viewed from the head, and, when using multiple prescriptions with forced densities.",
      "applies": "this device",
      "via": "K151233"
    }
  ],
  "eu_market_availability": null,
  "properties": [
    {
      "group": "Identifiers",
      "fields": [
        {
          "label": "510(k) Number",
          "values": [
            {
              "value": "K151233",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        },
        {
          "label": "FDA product code",
          "values": [
            {
              "value": "MUJ",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        }
      ]
    },
    {
      "group": "Classification",
      "fields": [
        {
          "label": "Device class",
          "values": [
            {
              "value": "Class II",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        },
        {
          "label": "Regulation",
          "values": [
            {
              "value": "21 CFR Part 892.5050",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        },
        {
          "label": "Advisory committee",
          "values": [
            {
              "value": "Radiology",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        },
        {
          "label": "FDA decision",
          "values": [
            {
              "value": "SESE",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        }
      ]
    }
  ]
}