Monaco RTP System
510(k) K183037
Active
Valid until
No expiry
FDA recalls
5 ongoingDocuments
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK183037 | Date4 Dec 2018 | StatusActive |
Recalls
-
- Ongoing Z-2563-2021
Contour changes can be saved on an unintended image set. In addition, these contour edits do not cause the dose to be frozen on the plans associated with that image set.
-
- Ongoing Z-2406-2021
If a couch is removed while a plan is loaded, and then added back to the Studyset when the plan is unloaded, it is possible to calculate dose with the couch unassigned to the beams.
-
- Ongoing Z-1825-2020
The Monaco RTP Radiation Treatment Planning System may change the shape and volume of the contour potentially resulting in the device delivering an inaccurate dose.
-
- Ongoing Z-0670-2020
Monaco is using the incorrect energy when optimizing and calculating dose.
-
- Ongoing Z-2560-2019
It is possible that the forced electron density settings will be changed for some structures unintentionally and this can result in incorrect dose calculation.
-
- Terminated Z-0965-2019
If Improve Target Dose was chosen as an optimization model in a previous treatment session, Monaco will automatically use this optimization model again when proceeding with the online plan adaptation of a completion plan when it should not.
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA Product Code
- MUJ
- Device Class
- Class II
- Regulation
- 21 CFR Part 892.5050
- Advisory Committee
- Radiology
- FDA Decision
- SESE Substantially Equivalent