- **510(k)** · FDA 510(k) · United States · K190178 · decision 2019-03-29 · decision SESE · status: Active

## Recalls

- Z-2563-2021 · Class II · ongoing · initiated 2021-09-14: Contour changes can be saved on an unintended image set. In addition, these contour edits do not cause the dose to be frozen on the plans associated with that image set.
- Z-2406-2021 · Class II · ongoing · initiated 2021-07-26: If a couch is removed while a plan is loaded, and then added back to the Studyset when the plan is unloaded, it is possible to calculate dose with the couch unassigned to the beams.
- Z-0670-2020 · Class II · ongoing · initiated 2019-11-22: Monaco is using the incorrect energy when optimizing and calculating dose.
- Z-2561-2019 · Class II · ongoing · initiated 2019-09-04: It is possible that the forced electron density settings will be changed for some structures unintentionally and this can result in incorrect dose calculation.
- Z-2409-2019 · Class II · terminated · initiated 2019-08-23: Monaco is displaying the Anatomy and Beam shift direction incorrectly on the Monaco Scan and Setup Reference Report and is DICOM exporting incorrect (Anatomy/Beam) shift directions.

## Identifiers

- 510(k) Number: K190178 [FDA 510(k)]
- FDA product code: MUJ [FDA 510(k)]

## Classification

- Device class: Class II [FDA 510(k)]
- Regulation: 21 CFR Part 892.5050 [FDA 510(k)]
- Advisory committee: Radiology [FDA 510(k)]
- FDA decision: SESE [FDA 510(k)]

## Source and citation

- Cite as: Monaco RTP System, Elekta, Inc. deviCERTS medical device catalogue. https://devicerts.com/devices/monaco-rtp-system-muj-k190178 (data as of 2026-11-20).
- Data sources: [FDA 510(k)](https://open.fda.gov/apis/device/510k/), [FDA recalls](https://open.fda.gov/apis/device/recall/)
- Registry records last synced: 2026-10-22
- Statuses computed as of: 2026-11-20
- JSON: <https://devicerts.com/devices/monaco-rtp-system-muj-k190178.json>
- Licence: [CC BY 4.0](https://creativecommons.org/licenses/by/4.0/). The deviCERTS layout and text are licensed under CC BY 4.0. Registry data remain under their source terms: U.S. FDA data (openFDA, AccessGUDID) are U.S. public domain; EUDAMED data are subject to the European Commission reuse terms.
- Attribution: DevicErts, https://devicerts.com/devices/monaco-rtp-system-muj-k190178
- Compiled by deviCERTS (devicerts.com) from public medical device registries (EUDAMED, European Commission; U.S. FDA via openFDA; AccessGUDID, U.S. National Library of Medicine). deviCERTS is not an official register and gives no regulatory advice: the record in the registry itself prevails.