MONARCH™ Planning Laptop Software P70 Linux USB Image Version 2.2.6
Listed
Valid until
Not applicable
FDA recalls
2 ongoingDocuments
None
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | GUDID listing | Number10810068811749 | Date12 Jun 2026 | StatusListed |
Recalls
-
- Ongoing Z-1383-2026
Device for bronchoscopic visualization, patient airway access has software issue: if application restarts after patient-side selection, prior to bronchoscope loading, application will re-initialize to patient-left position; and if right position was previously selected, this may lead to unanticipated robotic arm positioning toward patient-left, may result in contact with patient/endotracheal tube.
-
- Ongoing Z-0051-2025
Endoscopy/bronchoscopy/urology devices may have a strain relief defect with some cart and tower power cords that may cause an electrical short, which may lead to system shutdown, which can lead to prolonged procedure or case conversion, and if the user (e.g., operating room staff) touches exposed wires or conductive fluids in contact with the exposed wires, there is a chance of electrical shock.
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — GUDID listing Listed
Specification
- Trade Name
- MONARCH™ Planning Laptop Software P70 Linux USB Image
- Generic Name
- Robotic surgical telemanipulation system
- Model / Reference
- Version 2.2.6
- Primary DI / UDI-DI
- 10810068811749
- 510(k) Number
- FDA Product Code
- EOQ, JAK
- GMDN Term
- Robotic surgical telemanipulation system
- Device Class
- Class II
- Regulation Number
- 874.4680, 892.1750
- GUDID distribution status
- In Commercial Distribution
Other devices from Auris Health, Inc.
Properties
Other properties (13)
- Sterile: No
- Single Use: No
- Implantable: No
- Contains Latex: No
- Over-the-Counter: No
- Convenience Kit: No
- Combination Product: No
- HCT/P: No
- Premarket Exempt: No
- Lot / Batch Number: No
- Serial Number: No
- Expiration Date: No
- Manufacturing Date: No