Most Most 20-2B
by Pacific Consolidated Industries LLC United States GUDID
Product code CAW
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | GUDID listing | Number00860007698231 | Date8 Jun 2023 | StatusIn distribution |
Coverage
- Listed
- Not covered
United States — GUDID listing Listed
Recalls
-
- Terminated Z-2201-2008
The recall was initiated after the firm received two customer complaints of ruptured bourdon tubes that had resulted in bellowed out face gauges. With one unit the operator was able to shut off the oxygen flow, however one rupture resulted in a fire. No injuries were reported.
Specification
- Trade Name
- Most
- Generic Name
- Bulk oxygen concentration system
- Model / Ref / Catalog
- MOST 20-2B, 793080-002
- Primary DI / UDI-DI
- 00860007698231
- 510(k) Number
- FDA Product Code
- CAW
- GMDN Term
- Bulk oxygen concentration system
- Device Class
- Class II
- Regulation Number
- 868.5440
- Commercial Distribution
- In Commercial Distribution
Other devices from Pacific Consolidated Industries LLC
Properties & raw data
Show properties, registry sources and raw records
Other properties (11)
- Sterile: No
- Single Use: No
- Implantable: No
- Contains Latex: No
- Over-the-Counter: No
- Convenience Kit: No
- Combination Product: No
- HCT/P: No
- Premarket Exempt: No
- Lot / Batch Number: No
- Expiration Date: No
Sources (1)
- FDA UDI b3d7ea15-a75b-4188-a99f-b540cb1713aa