Multidiagnost Eleva
by Philips Medical Systems North America, Inc.
510(k) K050151
Active
Valid until
No expiry
FDA recalls
4 ongoingDocuments
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK050151 | Date9 Feb 2005 | StatusActive |
Recalls
-
- Ongoing Z-2411-2026
Philips Allura Xper and MultiDiagnost-Eleva systems Hand switch button not fully releasing potentially leading to loss of imaging functionality or unintended radiation exposure and additional contrast injection
-
- Ongoing Z-3168-2024
Potential Detachment of the Cable Hose Carrier (due to broken plastic and/or bolts) which May Result in Parts Falling and/or Part of the Cable Hose Dropping, and result in injury.
-
- Ongoing Z-2587-2023
A damaged foot switch may cause instances where no or only intermittent x-ray radiation initiation is possible.
-
- Ongoing Z-2404-2023
A foot switch pedal may get stuck in the active position when the user releases the pedal, resulting in the emission of unintended radiation
-
- Terminated Z-1489-2020
A capacitor inside the converter of the Velara X-ray generator may fail after a large number of surges in a short period of time. When this occurs, no image acquisition is possible.
-
- Terminated Z-1055-2019
The locking plate may be missing from the system, which may result in unexpected stand/table tilt movement that cannot be stopped by the user.
-
- Terminated Z-1140-2019
- Terminated Z-1141-2019
Additional units have been identified for previous recall Z-1280-2016 (RES 69339): If a Footswitch is frequently used on a non-flat surface or in the pedestal, the footswitch pedals may get bent. This might cause an intermittent or continuous inability of making live fluoroscopy images or exposures.
-
- Terminated Z-1609-2018
- Terminated Z-1606-2018
- Terminated Z-1607-2018
- Terminated Z-1608-2018
- Terminated Z-1603-2018
The fixation of the upper and lower tilt actuator, of the MD-Eleva system might break off and the table will start to rotate from 0 to + 90 I- 90 degree with high speed. This rotating movement cannot be stopped by the user.
-
- Terminated Z-1280-2016
- Terminated Z-1279-2016
Loss of key image functionality due to a bent pedal of the Footswitch.
-
- Terminated Z-1045-2015
The possibility of the Reference Air Kerma Rate for low dose mode will not be reduced to 50% of the value for the normal mode, but will be approximately 70-80%, leading to a higher patient dose than expected.
-
- Terminated Z-2156-2012
When doing a 3 Dimensional Rotational Angiography (3DRA) the 3D reconstruction does not always start, even when all images have been acquired.
-
- Terminated Z-1829-2012
MultiDiagnost Eleva X--Ray system may have damage to System Cables from repeated movement
-
- Terminated Z-0936-2012
The skin dose and skin dose rate values are calculated using an incorrect point of reference. Therefore, the AKR as displayed and stored in patient records is too low, although the dose as received by the patient is not affected.
-
- Terminated Z-0623-2009
System fails to meet requirements when it is used at kV levels greater than 115 kV at certain frame rates and in certain acquisition modes, as grid switch leakage may occur.
-
- Terminated Z-1137-2008
Delay in Imaging: MultiDiagnost Eleva with Flat Detector x-ray system cannot perform fluoroscopy procedures while certain image file housekeeping operations are performed. Fluoroscopy cannot be performed and new exposures are not able to be acquired until the housekeeping operations are complete which may take several minutes. The loss of system availability could occur during an interventional
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA Product Code
- KPR
- Device Class
- Class II
- Regulation
- 21 CFR Part 892.1680
- Advisory Committee
- Radiology
- FDA Decision
- SESE Substantially Equivalent