  "format": "devicerts.device.v1",
  "cite_as": "MX8000 V5.0 CT System, Philips Medical Systems (Cleveland), Inc. deviCERTS medical device catalogue. https://devicerts.com/devices/mx8000-v5-0-ct-system-jak (data as of 2026-11-22).",
  "source": {
    "canonical_url": "https://devicerts.com/devices/mx8000-v5-0-ct-system-jak",
    "json_url": "https://devicerts.com/devices/mx8000-v5-0-ct-system-jak.json",
    "markdown_url": "https://devicerts.com/devices/mx8000-v5-0-ct-system-jak.md",
    "publisher": "deviCERTS (devicerts.com)",
    "data_sources": [
      {
        "registry": "FDA 510(k)",
        "url": "https://open.fda.gov/apis/device/510k/"
      },
      {
        "registry": "FDA recalls",
        "url": "https://open.fda.gov/apis/device/recall/"
      }
    ],
    "registry_records_last_synced": "2026-10-22",
    "status_as_of": "2026-11-22",
    "license": "CC BY 4.0",
    "license_url": "https://creativecommons.org/licenses/by/4.0/",
    "license_scope": "The deviCERTS layout and text are licensed under CC BY 4.0. Registry data remain under their source terms: U.S. FDA data (openFDA, AccessGUDID) are U.S. public domain; EUDAMED data are subject to the European Commission reuse terms.",
    "attribution": "DevicErts, https://devicerts.com/devices/mx8000-v5-0-ct-system-jak",
    "disclaimer": "Compiled by deviCERTS (devicerts.com) from public medical device registries (EUDAMED, European Commission; U.S. FDA via openFDA; AccessGUDID, U.S. National Library of Medicine). deviCERTS is not an official register and gives no regulatory advice: the record in the registry itself prevails."
  },
  "device": {
    "name": "MX8000 V5.0 CT System",
    "registry_name": "MX8000 V5.0 CT SYSTEM",
    "url": "https://devicerts.com/devices/mx8000-v5-0-ct-system-jak",
    "manufacturer": {
      "name": "Philips Medical Systems (Cleveland), Inc.",
      "url": "https://devicerts.com/manufacturers/philips-medical-systems-cleveland-inc",
      "country": "United States",
      "country_basis": null,
      "fda_registration_country": null
    },
    "status": {
      "state": "active",
      "label": "Active",
      "detail": "United States"
    },
    "classification": [
      {
        "class": "Class II",
        "system": "US FDA"
      }
    ],
    "identifiers": [
      {
        "label": "510(k) Number",
        "values": [
          {
            "value": "K010817",
            "registries": [
              "FDA 510(k)"
            ]
          }
        ]
      }
    ],
    "product_codes": [
      "JAK"
    ],
    "emdn": [],
    "gmdn": [],
    "description": null,
    "noindex": false
  },
  "authorizations": [
    {
      "registry": "510(k)",
      "region": "United States",
      "type": "FDA 510(k)",
      "number": "K010817",
      "number_label": null,
      "device_udi_di": null,
      "date": "2001-06-15",
      "date_kind": "decision",
      "decision_code": "SESE",
      "expires": null,
      "withdrawn": null,
      "status": "Active",
      "status_note": null,
      "derived_from_gudid": false
    }
  ],
  "no_longer_listed_by": [],
  "certificates": [],
  "recalls": [
    {
      "number": "Z-1979-2021",
      "classification": "Class II",
      "status": "ongoing",
      "initiated": "2021-05-01",
      "reason": "Significant wear of the Monitor Bracket Assembly shaft  goes unrecognized could potentially lead to the Monitor Bracket Assembly detaching from the Suspension Arm and result in potential hazard for the monitor to fall while remaining attached to the\ncables resulting in patient, operator, bystander, or service person  to serious injury.",
      "applies": "this device",
      "via": "K010817"
    },
    {
      "number": "Z-1647-2015",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2015-03-16",
      "reason": "The firm was informed that while raising the patient couch on the system to perform an exam, the couch unexpectedly descended to the lowest point without being commanded to do so.",
      "applies": "this device",
      "via": "K010817"
    },
    {
      "number": "Z-1722-2014",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2014-04-01",
      "reason": "A customer reported that the table top had become free floating.  A Field Service Engineer (FSE) confirmed that the service latch had become disengaged.",
      "applies": "this device",
      "via": "K010817"
    },
    {
      "number": "Z-0945-2013",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2012-12-20",
      "reason": "Philips Healthcare received reports from the field stating the patient table on a system had an unexpected downward movement while a patient was on the table. Philips has determined the cause was a malfunction of the vertical brake.",
      "applies": "this device",
      "via": "K010817"
    },
    {
      "number": "Z-2914-2011",
      "classification": "",
      "status": "terminated",
      "initiated": "2011-06-03",
      "reason": "Philips Medical Systems decided to conduct a recall/field correction of the patient support vertical brake hub component in all of their CT MX8000 Dual v. Exp Computed Tomography X-Ray System scanners due to a component defect which could cause an unexpected downward motion of the table (couch) while the patient is on the table. This event could occur because the short key connecting the vertical",
      "applies": "shared submission",
      "via": "K010817"
    },
    {
      "number": "Z-0597-2009",
      "classification": "",
      "status": "terminated",
      "initiated": "2008-09-23",
      "reason": "The product is being recalled because the patient support may travel downwards without being commanded to move.",
      "applies": "shared submission",
      "via": "K010817"
    },
    {
      "number": "Z-0014-2007",
      "classification": "",
      "status": "terminated",
      "initiated": "2006-11-06",
      "reason": "A potential anomaly was identified on the CT subsytem, which could result in the radiation dose being different than the calculated dose reported to the operator.",
      "applies": "shared submission",
      "via": "K010817"
    },
    {
      "number": "Z-2709-2014",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2005-12-27",
      "reason": "The system has a gap that can appear between the upper and lower tilt stand covers. The fingers may be squeezed if a person was to place them in this region while the gantry was tilting from an angle to zero tilt angle.",
      "applies": "this device",
      "via": "K010817"
    },
    {
      "number": "Z-1412-05",
      "classification": "",
      "status": "terminated",
      "initiated": "2005-03-11",
      "reason": "When two product applications called Flip and Secondary Capture are used together, it is possible for image orientation labels to be incorrect.",
      "applies": "this device",
      "via": "K010817"
    },
    {
      "number": "Z-1414-05",
      "classification": "",
      "status": "terminated",
      "initiated": "2005-03-11",
      "reason": "When two product applications called Flip and Secondary Capture are used together, it is possible for image orientation labels to be incorrect.",
      "applies": "this device",
      "via": "K010817"
    },
    {
      "number": "Z-1410-05",
      "classification": "",
      "status": "terminated",
      "initiated": "2005-03-11",
      "reason": "When two product applications called Flip and Secondary Capture are used together, it is possible for image orientation labels to be incorrect.",
      "applies": "this device",
      "via": "K010817"
    },
    {
      "number": "Z-0836-03",
      "classification": "",
      "status": "terminated",
      "initiated": "2003-04-02",
      "reason": "Patient images may be incorrectly stored in the archive directory of a different patient.",
      "applies": "this device",
      "via": "K010817"
    }
  ],
  "eu_market_availability": null,
  "properties": [
    {
      "group": "Identifiers",
      "fields": [
        {
          "label": "510(k) Number",
          "values": [
            {
              "value": "K010817",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        },
        {
          "label": "FDA product code",
          "values": [
            {
              "value": "JAK",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        }
      ]
    },
    {
      "group": "Classification",
      "fields": [
        {
          "label": "Device class",
          "values": [
            {
              "value": "Class II",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        },
        {
          "label": "Regulation",
          "values": [
            {
              "value": "21 CFR Part 892.1750",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        },
        {
          "label": "Advisory committee",
          "values": [
            {
              "value": "Radiology",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        },
        {
          "label": "FDA decision",
          "values": [
            {
              "value": "SESE",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        }
      ]
    }
  ]
}