Recalls
-
- Terminated Z-1250-2014
A complaint was received for a discolored All Poly Tibia. Root cause analysis found that the discoloration was the result of free radical oxidation. The oxidation process was advanced prematurely as a result of the machined component being stored in a transparent container exposed to artificial light for six years prior to inert packaging and sterilization of the device.
-
- Terminated Z-0996-2009
The package contains 11 mm screws instead of the specified 6 mm screws.
-
- Terminated Z-0539-05
Risk of adapter disassociation from mating components when assembled improperly or when subjected to high-stress femoral applications.
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — GUDID listing Listed
Specification
- Generic Name
- Uncoated knee tibia prosthesis, metallic
- Model / Ref / Catalog
- 5020-03-012
- Primary DI / UDI-DI
- 00889024247918
- 510(k) Number
- FDA Product Code
- LZO, KRO
- GMDN Term
- Uncoated knee tibia prosthesis, metallic
- Device Class
- Class II
- Regulation Number
- 888.3353, 888.3510
- GUDID distribution status
- In Commercial Distribution
- Size
- Height — 12 Millimeter
Other devices from Zimmer, Inc.
Properties
Other properties (8)
- Contains Latex: No
- Over-the-Counter: No
- Convenience Kit: No
- Combination Product: No
- HCT/P: No
- Premarket Exempt: No
- Serial Number: No
- Manufacturing Date: No