8880T2 is a Class III device made by Medtronic, Inc., approved under FDA PMA P960009 on 31 Jul 1997, listed under FDA HDE H020007 (1 earlier submission) and listed in FDA GUDID since 2 Nov 2018 (UDI-DI 00763000017606), active as of 23 Oct 2026. 20 FDA recalls are linked to it.

Active United States. Also listed in GUDID (US).

PMA HDE

Valid until

No date given

No validity date on record

FDA recalls

8 ongoing

This device is named in the recall

Documents

PMA P960009

+2 more

Trust

Unverified

Source GUDID, synced 23 Oct 2026.

Regulatory status

as of registry sync 23 Oct 2026

RegionAuthorizationNumberDateStatus
United StatesGUDID listing Number00763000017606 UDI-DI Date2 Nov 2018ListedStatusListed GUDID: in commercial distribution
United StatesFDA PMAAPPR from GUDID Approved NumberP960009 Date31 Jul 1997DecidedStatusActive
United StatesFDA HDE from GUDID NumberH020007 Date—StatusActive

Official records

Across registries

3 registries list this device

  • FDA PMA P960009 decided 31 Jul 1997 Active
  • FDA HDE H020007 Active· +1 more in the status table
  • FDA GUDID UDI-DI 00763000017606 listed 2 Nov 2018 Listed

Recalls

20 recalls · 8 ongoing · FDA

  • 2 recalls, same reason

    • Ongoing Z-2240-2026 FDA Class II initiated 13 May 2026
    • Ongoing Z-2241-2026 FDA Class II initiated 13 May 2026

    A limited number of Pocket Adaptor Kits, Model 64001 and 64002, are labeled with incorrect Use-By-Dates.

    • Ongoing Z-0206-2025 FDA Class II initiated 24 Sep 2024

    A limited number of Percept RC Implantable Neurostimulators (INS), Model B35300 will not be able to communicate with the A610 DBS Clinician Programmer Application upon use and an "Invalid Device" message will appear.

    • Ongoing Z-0372-2025 FDA Class II initiated 18 Sep 2024

    The Wireless Rechargers in the kits may be unable to enter a recharging session upon first use.

    • Ongoing Z-1128-2022 FDA Class II initiated 30 Mar 2022

    The Implantable Neurostimulator (INS) cannot communicate with the clinical programmer and/or the patient programmer system.

    • Ongoing Z-0418-2022 FDA Class II initiated 19 Oct 2021

    During cardioversion events, the Model B35200 Percept PC Implantable Neurostimulator could become unresponsive and non-functional. Cardioversion is a procedure performed by a cardiologist to correct an abnormal heartbeat. Medtronic has determined that the cardioversion may damage the circuit electronics in the Percept PC stimulator, thereby, causing the Percept PC stimulator to become non-responsive.

    • Ongoing Z-0201-2022 FDA Class II initiated 23 Sep 2021

    A software anomaly may occur with the clinician programmer application.

    • Ongoing Z-1368-2021 FDA Class II initiated 9 Mar 2021

    There are two issues within this correction action: 1. Cycling Issue: When copying settings from a previously used Activa device to a new Percept PC Implantable Neurostimulator (INS) using the "replacement" function of the A610 clinician programmer application, the cycling feature is not copied properly. 2. Log Fill Issue: When the Percept PC INS System Event log is full, the clinician programmer A610 Software application version 2.0.4584, 2.0.4594, 2.0.4605, or 3.0.1057 crashes when it attempts a reading of the last records in the log.

    • Completed Z-0217-2024 FDA Class II initiated 27 May 2022

    There are a small number of units of the DBS Lead with StimLoc kit (3387S-40) due to a discrepancy in the Use By Date (UBD) printed on the outer packaged kit label versus the UBD printed on the individual product label.

    • Terminated Z-3069-2017 FDA Class II initiated 17 Mar 2017

    Medtronic received reports that the DBS depth stop did not adequately secure to the lead, which can result in DBS lead placement beyond the intended target.

    • Terminated Z-2645-2016 FDA Class II initiated 23 Jun 2016

    Due to the potential for pocket adaptor conductor wire fractures involving DBS Pocket Adaptor Model 64001 and Model 64002, a safety notification was sent to healthcare professionals to reinforce device specific labeling for the handling of pocket adaptors and system integrity checking during implant procedures.

    • Terminated Z-2602-2016 FDA Class II initiated 10 Jun 2016

    Medtronic has voluntarily decided to retrieve unused Model 3387S-40 and 3389S-40 DBS leads from three manufacturing lot numbers. During the manufacturing process of a DBS lead component, Medtronic identified the potential for lead insulation damage.

    • Terminated Z-1617-2015 FDA Class II initiated 16 Apr 2015

    Medtronic is providing Important information regarding extension handling during implant procedure after receiving reports of high impedances. This applies to extension Models 37085 and 37086 which can be used with the following implantable Neurostimulators: Activa PC (Model 37601), Activa RC (Model 37612), and the Activa SC (Model 37603).

    • Terminated Z-0279-2014 FDA Class II initiated 30 Sep 2013

    Medtronic notified customers regarding Loss of Stimulation of specific Implantable Neuro-Stimulators under the following specific conditions: Switching between groups with Multiple programs or creating program groups. Therapy can be restored with the Physician Recharge mode (RPM) of the Implantable Neurostimulator Recharger (INSR).

    • Terminated Z-1190-2013 FDA Class I initiated 8 Feb 2013

    There is a potential for lead damage due to the use of the lead cap provided in DBS and Dystonia kits. Medtronic has received reports of DBS leads being damaged at the connector end of the lead when the lead cap is used. The connector end of the lead is the end of the lead connected to the lead extension. Tightening or loosening of the setscrew may twist the setscrew connector block and may damag

  • 4 recalls, same reason

    • Terminated Z-0184-2009 FDA Unclassified (pre-2012) initiated 14 Jun 2008
    • Terminated Z-0187-2009 FDA Unclassified (pre-2012) initiated 14 Jun 2008
    • Terminated Z-0186-2009 FDA Unclassified (pre-2012) initiated 14 Jun 2008
    • Terminated Z-0185-2009 FDA Unclassified (pre-2012) initiated 14 Jun 2008

    Medtronic has received reports of DBS" leads being damaged at the proximal connector end of the lead when the lead cap is used in the implant procedure. The lead damage is the result of excessive torsion and / or tensile lading of the proximal connector end of the lead. Data also suggest that excessive torque while tightening or loosening the setscrews may twist the setscrew connector block and d

    • Terminated Z-0370-06 FDA Unclassified (pre-2012) initiated 30 Nov 2005

    A safety concern exists regarding the Medtronic Activa Deep Brain Stimulation system instructions for performing MRI (magnetic resonance imaging). Medtronic's instructions have until now recommended a use of a displayed MRI head specific absorption rate (SAR) of not more than 0.4 W/kg. The recommended maximum displayed head SAR is now 0.1 W/kg.

Timeline

22 dated events

  1. Listed FDA GUDID 00763000017606
  2. Recall initiated FDA recalls Z-1368-2021
  3. Recall initiated FDA recalls Z-0201-2022
  4. Recall initiated FDA recalls Z-0418-2022
  5. Recall initiated FDA recalls Z-1128-2022
  6. Recall initiated FDA recalls Z-0217-2024
  7. Recall initiated FDA recalls Z-0372-2025
  8. Recall initiated FDA recalls Z-0206-2025
  9. Recall initiated FDA recalls Z-2240-2026
  10. Recall initiated FDA recalls Z-2241-2026

12 earlier events are in the sections above.

Coverage

1 jurisdiction

United States of America
  • Active
  • Expiring
  • Withdrawn
  • Listed
  • Not covered

United States — FDA HDE, FDA PMA Active

Specification

GUDID

Generic name
Multiple active implantable device programmer
Model / reference
8880T2
Primary DI / UDI-DI
00763000017606
FDA product code
LGW, MRU, OLM, NHL, MHY, MBX, LKK, PJS, QRB, SGS
GMDN term
Multiple active implantable device programmer
Device class
Class III