N/A PT101USR
by Carefusion
Listed
Valid until
Not applicable
FDA recalls
2 ongoingDocuments
None
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | GUDID listing | Number10190752102484 | Date4 Nov 2016 | StatusListed |
Recalls
-
- Ongoing Z-0158-2026
When unintentionally disconnected from power source, humidifier device (used to deliver high flow warmed and humidified respiratory gases) may sound audible alert for less than 120 seconds, which may delay user awareness to reconnect power, which could lead to oxygen desaturation, hypoxia. Disinfection kit user manual updated to include checking the audible alert prior to use on a new patient.
-
- Ongoing Z-1823-2024
Humidifier devices, used to deliver high flow respiratory therapy to patients, manufactured before 14 August 2017, have a speaker configuration that may result in distorted, intermittent or inaudible alarm sound levels. In the absence of an audible alarm, if there is an interruption to therapy, a patient may experience oxygen desaturation.
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — GUDID listing Listed
Specification
- Generic Name
- Heated respiratory humidifier
- Model / Ref / Catalog
- PT101USR
- Primary DI / UDI-DI
- 10190752102484
- 510(k) Number
- FDA Product Code
- BTT
- GMDN Term
- Heated respiratory humidifier
- Device Class
- Class II
- Regulation Number
- 868.5450
- GUDID distribution status
- Not in Commercial Distribution
Other devices from Carefusion
Properties
Other properties (12)
- Sterile: No
- Single Use: No
- Implantable: No
- Contains Latex: No
- Over-the-Counter: No
- Convenience Kit: No
- Combination Product: No
- HCT/P: No
- Premarket Exempt: No
- Lot / Batch Number: No
- Expiration Date: No
- Manufacturing Date: No