N Latex Cdt
by Dade Behring, Inc. 510(k)
Active
510(k) K060677
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK060677 | Date25 Aug 2006 | StatusActive |
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not known
- Not covered
United States — FDA 510(k) Active
Recalls
-
- Terminated Z-1790-2020
Negative bias for the carbohydrate-deficient transferrin (CDT) measurement in comparison to the HPLC method, observed to occur after 15 months based on shelf life monitoring
-
- Terminated Z-2931-2018
Siemens Healthcare Diagnostics has observed a negative bias for the carbohydrate-deficient transferrin (CDT) measurement when using the affected N Latex CDT lots in comparison to the HPLC method. CDT results in absolute concentrations may be influenced by patients transferrin levels and, therefore, results are reported as ratio of CDT to total transferrin, called %CDT. As the transferrin determination with N Antiserum to Human Transferrin is not affected, the observed effect leads to calculated %CDT values that show a negative bias of approximately 15% compared to the HPLC method. This could result in a shift of weak positive patient samples into the reference range of 1.19 2.47 %CDT which was derived from a study population of healthy adults. There are no control or calibrator materials that would catch this bias.
Specification
- 510(k) Number
- FDA Product Code
- NAO
- Device Class
- Class I
- Regulation
- 21 CFR Part 862.1360
- Advisory Committee
- Clinical Chemistry
- FDA Decision
- SESE Substantially Equivalent