Na 2347-22-30
by Zimmer, Inc.
Listed
Valid until
Not applicable
FDA recalls
None ongoingDocuments
None
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | GUDID listing | Number00889024052222 | Date16 Oct 2015 | StatusListed |
Recalls
-
- Terminated Z-0315-2020
- Terminated Z-0307-2020
Non-highly polished implants may adhere to the LDPE bag (previous packaging configuration).
-
- Terminated Z-1684-2016
A small hole may be present in the inner tray of specific lots of sterile Hip and Trauma Implant Screws.
-
- Terminated Z-1161-2016
- Terminated Z-1115-2016
- Terminated Z-1145-2016
- Terminated Z-1123-2016
As a result of the insufficient sealer calibration data for product packaged in firm's Building II between August 2010 and April 2013. The affected products are sterile.
-
- Terminated Z-0875-2015
Zimmer received a single complaint wherein a Trilogy® Fiber Metal Multi-Hole Acetabular Shell would not attach to the inserter/impactor instrumentation. Subsequent investigation found the polar-hole threads to be present but non-conforming to specifications, in that the minor diameter was undersized. Investigation also indicates that the issue is isolated and affects only lot 62714305, which
-
- Terminated Z-0824-2015
Analysis of returned complaint devices, product in inventory, and device manufacturing records has identified tool marks isolated to one specific machining mill to be the cause of the fractures. .
-
- Terminated Z-2032-2014
Two lots of screws were commingled. Screws with etching and machining for lot number 62629432 were packaged in bags labeled for lot number 62628759; and screws with etching and machining for lot number 62628759 were packaged in bags labeled for lot number 62629432. The affected product was distributed between 3/6/14 and 4/21/14.
-
- Terminated Z-1024-2014
Zimmer recently conducted a review of all historic packaging validations at its Zimmer Manufacturing B.V. (ZMBV) facility in Ponce, Puerto Rico. Based on this review, it was determined that the packaging operations conducted in the Building 1 packaging operation at this facility were not properly validated. As a result, Zimmer conducted seal integrity testing to evaluate if previously packaged product from this facility met the established requirements for sterile barrier integrity. The acceptance criteria established for this testing was 99% reliability with a 90% confidence interval. These packaging tray configurations met the acceptance criteria for sterile barrier integrity of both the inner and outer trays, however, Zimmer has decided to recall all manufacturing lots that have not expired and that were sterile packaged in the ZMBV facility Building 1 packaging operations.
-
- Terminated Z-0773-2013
- Terminated Z-0772-2013
- Terminated Z-0771-2013
- Terminated Z-0769-2013
- Terminated Z-0775-2013
Some units of Trilogy shells manufactured since March 2009 may be below the lower range of the pore size specification (as low as 28% vs. 30% porosity specification). On 10/31/2012, the firm expanded the recall action to include lots of Uni-Hole Trilogy Acetabular System shells.
-
- Terminated Z-0620-2008
Parts missing and incomplete etch: The units may not contain the locking ring or the etched alignment marks on the rim.
-
- Terminated Z-1344-06
Lack of assurance of sterility, as the sterile barrier packaging may have been compromised.
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — GUDID listing Listed
Specification
- Generic Name
- Orthopaedic bone screw (non-sliding)
- Model / Ref / Catalog
- 2347-22-30, 00-2347-022-30
- Primary DI / UDI-DI
- 00889024052222
- FDA Product Code
- HWC, LPH
- GMDN Term
- Orthopaedic bone screw (non-sliding)
- Device Class
- Class II
- Regulation Number
- 888.3040, 888.3358
- GUDID distribution status
- In Commercial Distribution
- Sterilization Method
- Moist Heat or Steam Sterilization
- Size
- Length — 30 Millimeter
Other devices from Zimmer, Inc.
Properties
Other properties (10)
- Sterile: No
- Contains Latex: No
- Over-the-Counter: No
- Convenience Kit: No
- Combination Product: No
- HCT/P: No
- Premarket Exempt: No
- Serial Number: No
- Expiration Date: No
- Manufacturing Date: No