00-2358-009-03
by Zimmer, Inc.
Ref 00-2358-009-03
00-2358-009-03 is an orthopaedic fixation plate (EU MDR Class IIb; FDA Class II; FDA product code HRS, Plate, Fixation, Bone) made by Zimmer, Inc., registered in EUDAMED on 2 Nov 2026 (Basic UDI-DI 889024-ZB001086-FK), cleared under FDA 510(k) K141734 on 13 Aug 2014 (1 earlier submission) and listed in FDA GUDID since 9 Oct 2015 (UDI-DI 00889024056312), active as of 7 Nov 2026. 7 FDA recalls are linked to it.
Active
Valid until
No date given
FDA recalls
1 ongoingTrust
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| European Union | CE/MDR EUDAMED | Number889024-ZB001086-FK 00889024056312 | Date2 Nov 2026 | StatusActive |
| United States | GUDID listing GUDID | Number00889024056312 | Date9 Oct 2015 | StatusListed |
| United States | FDA 510(k)SESE GUDID | NumberK141734 | Date13 Aug 2014 | StatusActive |
Official records
- EUDAMED public site 889024-ZB001086-FK
- AccessGUDID record 00889024056312
- FDA 510(k) database K141734
- FDA 510(k) database K082078
From the FDA record
- MRI safety
- Labeling does not contain MRI Safety Information
- Natural rubber latex
- No latex statement on the label
- Sterility
- Supplied non-sterile; sterilise before use (Moist Heat or Steam Sterilization)
Across registries
- EUDAMED Basic UDI-DI 889024-ZB001086-FK registered 2 Nov 2026 Active
- FDA 510(k) K141734 decided 13 Aug 2014 Active
- FDA GUDID UDI-DI 00889024056312 listed 9 Oct 2015 Listed
Recalls
-
- Ongoing Z-2254-2023
Thread form issue of the locking holes in that the locking screws would not properly mate with the plate. The improperly mated screw may not be readily recognizable by the user since the screw may not correctly lock.
-
- Terminated Z-0310-2020
- Terminated Z-0317-2020
Non-highly polished implants may adhere to the LDPE bag (previous packaging configuration).
-
- Terminated Z-0585-2014
- Terminated Z-0587-2014
- Terminated Z-0588-2014
- Terminated Z-0589-2014
- Terminated Z-0591-2014
Based on a review of historic packaging validations, it was determined that a subset of product packaged at a Zimmer facility required testing to ensure that previously packaged product from this facility met the established requirements for sterile barrier integrity. This testing determined that a packaging tray configuration which consists of an inner and outer sterile tray, resulted in failure
-
- Terminated Z-2101-2013
Zimmer is initiating a lot specific voluntary recall of the Distal Medial Humeral Locking Plate, Short due to the incorrect information printed on the Chinese label. The original label indicates that implants are RIGHT sided, but the Chinese label indicates that the implants are LEFT sided.
Timeline
- Recall initiated FDA recalls Z-0587-2014
- Recall initiated FDA recalls Z-0588-2014
- Recall initiated FDA recalls Z-0589-2014
- Recall initiated FDA recalls Z-0591-2014
- Cleared FDA 510(k) K141734
- Listed FDA GUDID 00889024056312
- Recall initiated FDA recalls Z-0310-2020
- Recall initiated FDA recalls Z-0317-2020
- Recall initiated FDA recalls Z-2254-2023
- Registered EUDAMED 889024-ZB001086-FK
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
- European Union
- United States
Specification
- Trade name
- Zimmer® Periarticular Locking Plate System
- Generic name
- Orthopaedic fixation plate, non-bioabsorbable
- Device name
- ZPLP Plate Ulna Proximal Right
- Model / Ref / Catalog
- 00-2358-009-03
- Primary DI / UDI-DI
- 00889024056312
- Basic UDI-DI
- 889024-ZB001086-FK
- FDA product code
- HRS
- EMDN code
- P09120503 Osteosynthesis screw-plate systems
- GMDN term
- Orthopaedic fixation plate, non-bioabsorbable
- Device class
- Class IIb EUDAMEDClass II GUDID
- Regulation number
- 888.3030
- EU registration type
- Standard device
- EU market of registration
- Netherlands
- GUDID distribution status
- In Commercial Distribution
- Sterilisation method
- Moist Heat or Steam Sterilization
- Size
- Length — 77 Millimeter
Truth per registry
| Field | EUDAMED | GUDID |
|---|---|---|
| Device class | Class IIb | Class II |
Similar devices in this category
Other devices filed under OSTEOSYNTHESIS SCREW-PLATE SYSTEMS.
- 2.0MM SYSTEM SP-2626 by BIOMET MICROFIXATION, INC
- A.L.P.S.® Elbow System 1312-18-606 131218606 by Biomet Orthopedics, Inc.
- Arthrex® AR-9963PLL-10S by Arthrex, Inc.
- Arthrex® AR-8952XL by Arthrex, Inc.
- LOQTEQ® PU 3532-06-2 by aap Implantate AG
- A.L.P.S.® 824077024 by Biomet Orthopedics, Inc.
- NEOFUSE V45 ST185 by In2Bones SAS
- NEOFUSE V45 ST524 by In2Bones SAS
- Orthomed E® Bone Fixation Plates 6222049604313 by Orthomed E
- Orthomed E® Bone Fixation Plates 6222065349250 by Orthomed E