00-2358-107-07
by Zimmer, Inc.
Ref 00-2358-107-07
00-2358-107-07 is an orthopaedic fixation plate (EU MDR Class IIb; FDA Class II; FDA product code HRS, Plate, fixation, bone) made by Zimmer, Inc., registered in EUDAMED on 2 Nov 2026 (Basic UDI-DI 889024-ZB001052-EQ), cleared under FDA 510(k) K082078 on 3 Nov 2008 and listed in FDA GUDID since 17 Oct 2015 (UDI-DI 00889024057210), active as of 7 Nov 2026. 5 FDA recalls are linked to it.
Active
Valid until
No date given
FDA recalls
None ongoingTrust
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| European Union | CE/MDR EUDAMED | Number889024-ZB001052-EQ 00889024057210 | Date2 Nov 2026 | StatusActive |
| United States | GUDID listing GUDID | Number00889024057210 | Date17 Oct 2015 | StatusListed |
| United States | FDA 510(k)SESE GUDID | NumberK082078 | Date3 Nov 2008 | StatusActive |
Official records
- EUDAMED public site 889024-ZB001052-EQ
- AccessGUDID record 00889024057210
- FDA 510(k) database K082078
From the FDA record
- MRI safety
- Labeling does not contain MRI Safety Information
- Natural rubber latex
- No latex statement on the label
- Sterility
- Supplied non-sterile; sterilise before use (Moist Heat or Steam Sterilization)
Across registries
- EUDAMED Basic UDI-DI 889024-ZB001052-EQ registered 2 Nov 2026 Active
- FDA 510(k) K082078 decided 3 Nov 2008 Active
- FDA GUDID UDI-DI 00889024057210 listed 17 Oct 2015 Listed
Recalls
-
- Terminated Z-0317-2020
Non-highly polished implants may adhere to the LDPE bag (previous packaging configuration).
-
- Terminated Z-0585-2014
- Terminated Z-0587-2014
- Terminated Z-0588-2014
- Terminated Z-0589-2014
- Terminated Z-0591-2014
Based on a review of historic packaging validations, it was determined that a subset of product packaged at a Zimmer facility required testing to ensure that previously packaged product from this facility met the established requirements for sterile barrier integrity. This testing determined that a packaging tray configuration which consists of an inner and outer sterile tray, resulted in failure
-
- Terminated Z-2101-2013
Zimmer is initiating a lot specific voluntary recall of the Distal Medial Humeral Locking Plate, Short due to the incorrect information printed on the Chinese label. The original label indicates that implants are RIGHT sided, but the Chinese label indicates that the implants are LEFT sided.
Timeline
- Cleared FDA 510(k) K082078
- Recall initiated FDA recalls Z-2101-2013
- Recall initiated FDA recalls Z-0585-2014
- Recall initiated FDA recalls Z-0587-2014
- Recall initiated FDA recalls Z-0588-2014
- Recall initiated FDA recalls Z-0589-2014
- Recall initiated FDA recalls Z-0591-2014
- Listed FDA GUDID 00889024057210
- Recall initiated FDA recalls Z-0317-2020
- Registered EUDAMED 889024-ZB001052-EQ
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
- European Union
- United States
Specification
- Trade name
- Zimmer® Periarticular Locking Plate System
- Generic name
- Orthopaedic fixation plate, non-bioabsorbable
- Device name
- ZPLP Plate Humoral Distal Medial Right Short
- Model / Ref / Catalog
- 00-2358-107-07
- Primary DI / UDI-DI
- 00889024057210
- Basic UDI-DI
- 889024-ZB001052-EQ
- 510(k) Number
- FDA product code
- HRS
- EMDN code
- P09120503 Osteosynthesis screw-plate systems
- GMDN term
- Orthopaedic fixation plate, non-bioabsorbable
- Device class
- Class IIb EUDAMEDClass II GUDID
- Regulation number
- 888.3030
- EU registration type
- Standard device
- EU market of registration
- Netherlands
- GUDID distribution status
- In Commercial Distribution
- Sterilisation method
- Moist Heat or Steam Sterilization
- Size
- Length — 126 Millimeter
Truth per registry
| Field | EUDAMED | GUDID |
|---|---|---|
| Device class | Class IIb | Class II |
Similar devices in this category
Other devices filed under OSTEOSYNTHESIS SCREW-PLATE SYSTEMS.
- RIBFIX BLU SYSTEM 76-2604 by BIOMET MICROFIXATION, INC
- Orthomed E® Bone Fixation Plates 6222049664911 by Orthomed E
- Arthrex® AR-13715-17.5 by Arthrex, Inc.
- STERILETRAC SYSTEM 25-1020X-12 by BIOMET MICROFIXATION, INC
- Arthrex® AR-18724P-23 by Arthrex, Inc.
- A.L.P.S.® FOOT 8240-73-003 824073003 by Biomet Orthopedics, Inc.
- LOQTEQ® PH 3510-03-2 by aap Implantate AG
- LOQTEQ® PH 3541-11-2 by aap Implantate AG
- LOCKING DISTAL TIBIA ANTEROLATERAL PLATE - LEFT WIDE 8162-01-012 816201012 by Biomet Orthopedics, Inc.
- 181-10006 by TDM Co., Ltd.