  "format": "devicerts.device.v1",
  "cite_as": "9026-21, Zimmer, Inc. deviCERTS medical device catalogue. https://devicerts.com/devices/na-00-9026-021-00 (data as of 2026-11-20).",
  "source": {
    "canonical_url": "https://devicerts.com/devices/na-00-9026-021-00",
    "json_url": "https://devicerts.com/devices/na-00-9026-021-00.json",
    "markdown_url": "https://devicerts.com/devices/na-00-9026-021-00.md",
    "publisher": "deviCERTS (devicerts.com)",
    "data_sources": [
      {
        "registry": "FDA GUDID (AccessGUDID)",
        "url": "https://accessgudid.nlm.nih.gov/"
      },
      {
        "registry": "FDA recalls",
        "url": "https://open.fda.gov/apis/device/recall/"
      }
    ],
    "registry_records_last_synced": "2026-10-23",
    "status_as_of": "2026-11-20",
    "license": "CC BY 4.0",
    "license_url": "https://creativecommons.org/licenses/by/4.0/",
    "license_scope": "The deviCERTS layout and text are licensed under CC BY 4.0. Registry data remain under their source terms: U.S. FDA data (openFDA, AccessGUDID) are U.S. public domain; EUDAMED data are subject to the European Commission reuse terms.",
    "attribution": "DevicErts, https://devicerts.com/devices/na-00-9026-021-00",
    "disclaimer": "Compiled by deviCERTS (devicerts.com) from public medical device registries (EUDAMED, European Commission; U.S. FDA via openFDA; AccessGUDID, U.S. National Library of Medicine). deviCERTS is not an official register and gives no regulatory advice: the record in the registry itself prevails."
  },
  "device": {
    "name": "9026-21",
    "registry_name": "9026-21",
    "url": "https://devicerts.com/devices/na-00-9026-021-00",
    "manufacturer": {
      "name": "Zimmer, Inc.",
      "url": "https://devicerts.com/manufacturers/zimmer-inc",
      "country": "United States",
      "country_basis": null,
      "fda_registration_country": null
    },
    "status": {
      "state": "active",
      "label": "Active",
      "detail": "United States"
    },
    "classification": [
      {
        "class": "Class II",
        "system": "US FDA"
      }
    ],
    "identifiers": [
      {
        "label": "Catalogue number",
        "values": [
          {
            "value": "00-9026-021-00",
            "registries": [
              "FDA GUDID (AccessGUDID)"
            ]
          }
        ]
      },
      {
        "label": "Primary DI / UDI-DI",
        "values": [
          {
            "value": "00889024155978",
            "registries": [
              "FDA GUDID (AccessGUDID)"
            ]
          }
        ]
      },
      {
        "label": "510(k) Number",
        "values": [
          {
            "value": "K840643",
            "registries": [
              "FDA GUDID (AccessGUDID)"
            ]
          },
          {
            "value": "K953337",
            "registries": [
              "FDA GUDID (AccessGUDID)"
            ]
          }
        ]
      }
    ],
    "product_codes": [
      "LPH",
      "JDI"
    ],
    "emdn": [],
    "gmdn": [
      {
        "code": "33175",
        "term": "Metallic femoral head prosthesis"
      }
    ],
    "description": null,
    "noindex": true
  },
  "authorizations": [
    {
      "registry": "GUDID",
      "region": "United States",
      "type": "GUDID listing",
      "number": "00889024155978",
      "number_label": "UDI-DI",
      "device_udi_di": null,
      "date": "2015-09-24",
      "date_kind": "listed",
      "decision_code": null,
      "expires": null,
      "withdrawn": null,
      "status": "Listed",
      "status_note": "GUDID: in commercial distribution",
      "derived_from_gudid": false
    },
    {
      "registry": "GUDID",
      "region": "United States",
      "type": "FDA 510(k)",
      "number": "K953337",
      "number_label": null,
      "device_udi_di": null,
      "date": "1996-01-22",
      "date_kind": "decision",
      "decision_code": "SESE",
      "expires": null,
      "withdrawn": null,
      "status": "Active",
      "status_note": null,
      "derived_from_gudid": true
    },
    {
      "registry": "GUDID",
      "region": "United States",
      "type": "FDA 510(k)",
      "number": "K840643",
      "number_label": null,
      "device_udi_di": null,
      "date": "1984-04-17",
      "date_kind": "decision",
      "decision_code": "SESE",
      "expires": null,
      "withdrawn": null,
      "status": "Active",
      "status_note": null,
      "derived_from_gudid": true
    }
  ],
  "no_longer_listed_by": [],
  "certificates": [],
  "recalls": [
    {
      "number": "Z-0722-2018",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2017-08-01",
      "reason": "The products being recalled exceeded the weight previously tested leading to a possibility of damage to the package and/or a compromised sterile barrier.",
      "applies": "this device",
      "via": "K953337"
    },
    {
      "number": "Z-2580-2017",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2017-02-28",
      "reason": "Zimmer Biomet Inc. initiated a field action on 02/28/2017 for sterile-packaged implants packaged in two different package configurations due to packaging design verification test failures.",
      "applies": "this device",
      "via": "K953337"
    },
    {
      "number": "Z-2577-2017",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2017-02-28",
      "reason": "Zimmer Biomet Inc. initiated a field action on 02/28/2017 for sterile-packaged implants packaged in two different package configurations due to packaging design verification test failures.",
      "applies": "this device",
      "via": "K953337"
    },
    {
      "number": "Z-2581-2017",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2017-02-28",
      "reason": "Zimmer Biomet Inc. initiated a field action on 02/28/2017 for sterile-packaged implants packaged in two different package configurations due to packaging design verification test failures.",
      "applies": "this device",
      "via": "K953337"
    },
    {
      "number": "Z-2579-2017",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2017-02-28",
      "reason": "Zimmer Biomet Inc. initiated a field action on 02/28/2017 for sterile-packaged implants packaged in two different package configurations due to packaging design verification test failures.",
      "applies": "this device",
      "via": "K953337"
    },
    {
      "number": "Z-2578-2017",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2017-02-28",
      "reason": "Zimmer Biomet Inc. initiated a field action on 02/28/2017 for sterile-packaged implants packaged in two different package configurations due to packaging design verification test failures.",
      "applies": "this device",
      "via": "K953337"
    },
    {
      "number": "Z-1683-2016",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2016-05-04",
      "reason": "A small hole may be present in the inner tray of specific lots of sterile Hip and Trauma Implant Screws.",
      "applies": "this device",
      "via": "K840643"
    },
    {
      "number": "Z-1126-2016",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2016-01-11",
      "reason": "As a result of the insufficient sealer calibration data for product packaged in firm's Building II between August 2010 and April 2013.  The affected products are sterile.",
      "applies": "this device",
      "via": "K840643"
    },
    {
      "number": "Z-1137-2016",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2016-01-11",
      "reason": "As a result of the insufficient sealer calibration data for product packaged in firm's Building II between August 2010 and April 2013.  The affected products are sterile.",
      "applies": "this device",
      "via": "K840643"
    },
    {
      "number": "Z-1127-2016",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2016-01-11",
      "reason": "As a result of the insufficient sealer calibration data for product packaged in firm's Building II between August 2010 and April 2013.  The affected products are sterile.",
      "applies": "this device",
      "via": "K953337"
    },
    {
      "number": "Z-1131-2016",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2016-01-11",
      "reason": "As a result of the insufficient sealer calibration data for product packaged in firm's Building II between August 2010 and April 2013.  The affected products are sterile.",
      "applies": "this device",
      "via": "K840643"
    },
    {
      "number": "Z-1110-2016",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2016-01-11",
      "reason": "As a result of the insufficient sealer calibration data for product packaged in firm's Building II between August 2010 and April 2013.  The affected products are sterile.",
      "applies": "this device",
      "via": "K840643"
    },
    {
      "number": "Z-0849-2016",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2016-01-11",
      "reason": "LDPE bag containing the implant adheres to the highly polished implant surface.",
      "applies": "this device",
      "via": "K953337"
    },
    {
      "number": "Z-1033-2014",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2013-10-11",
      "reason": "Zimmer recently conducted a review of all historic packaging validations at its Zimmer Manufacturing B.V. (ZMBV) facility in Ponce, Puerto Rico. Based on this review, it was determined that the packaging operations conducted in the Building 1 packaging operation at this facility were not properly validated. As a result, Zimmer conducted seal integrity testing to evaluate if previously packaged product from this facility met the established requirements for sterile barrier integrity. The acceptance criteria established for this testing was 99% reliability with a 90% confidence interval. These packaging tray configurations met the acceptance criteria for sterile barrier integrity of both the inner and outer trays, however, Zimmer has decided to recall all manufacturing lots that have not expired and that were sterile packaged in the ZMBV facility Building 1 packaging operations.",
      "applies": "this device",
      "via": "K840643"
    },
    {
      "number": "Z-1034-2014",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2013-10-11",
      "reason": "Zimmer recently conducted a review of all historic packaging validations at its Zimmer Manufacturing B.V. (ZMBV) facility in Ponce, Puerto Rico. Based on this review, it was determined that the packaging operations conducted in the Building 1 packaging operation at this facility were not properly validated. As a result, Zimmer conducted seal integrity testing to evaluate if previously packaged product from this facility met the established requirements for sterile barrier integrity. The acceptance criteria established for this testing was 99% reliability with a 90% confidence interval. These packaging tray configurations met the acceptance criteria for sterile barrier integrity of both the inner and outer trays, however, Zimmer has decided to recall all manufacturing lots that have not expired and that were sterile packaged in the ZMBV facility Building 1 packaging operations.",
      "applies": "this device",
      "via": "K953337"
    },
    {
      "number": "Z-1025-2014",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2013-10-11",
      "reason": "Zimmer recently conducted a review of all historic packaging validations at its Zimmer Manufacturing B.V. (ZMBV) facility in Ponce, Puerto Rico. Based on this review, it was determined that the packaging operations conducted in the Building 1 packaging operation at this facility were not properly validated. As a result, Zimmer conducted seal integrity testing to evaluate if previously packaged product from this facility met the established requirements for sterile barrier integrity. The acceptance criteria established for this testing was 99% reliability with a 90% confidence interval. These packaging tray configurations met the acceptance criteria for sterile barrier integrity of both the inner and outer trays, however, Zimmer has decided to recall all manufacturing lots that have not expired and that were sterile packaged in the ZMBV facility Building 1 packaging operations.",
      "applies": "this device",
      "via": "K840643"
    },
    {
      "number": "Z-1031-2014",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2013-10-11",
      "reason": "Zimmer recently conducted a review of all historic packaging validations at its Zimmer Manufacturing B.V. (ZMBV) facility in Ponce, Puerto Rico. Based on this review, it was determined that the packaging operations conducted in the Building 1 packaging operation at this facility were not properly validated. As a result, Zimmer conducted seal integrity testing to evaluate if previously packaged product from this facility met the established requirements for sterile barrier integrity. The acceptance criteria established for this testing was 99% reliability with a 90% confidence interval. These packaging tray configurations met the acceptance criteria for sterile barrier integrity of both the inner and outer trays, however, Zimmer has decided to recall all manufacturing lots that have not expired and that were sterile packaged in the ZMBV facility Building 1 packaging operations.",
      "applies": "this device",
      "via": "K953337"
    },
    {
      "number": "Z-1036-2014",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2013-10-11",
      "reason": "Zimmer recently conducted a review of all historic packaging validations at its Zimmer Manufacturing B.V. (ZMBV) facility in Ponce, Puerto Rico. Based on this review, it was determined that the packaging operations conducted in the Building 1 packaging operation at this facility were not properly validated. As a result, Zimmer conducted seal integrity testing to evaluate if previously packaged product from this facility met the established requirements for sterile barrier integrity. The acceptance criteria established for this testing was 99% reliability with a 90% confidence interval. These packaging tray configurations met the acceptance criteria for sterile barrier integrity of both the inner and outer trays, however, Zimmer has decided to recall all manufacturing lots that have not expired and that were sterile packaged in the ZMBV facility Building 1 packaging operations.",
      "applies": "this device",
      "via": "K840643"
    },
    {
      "number": "Z-0070-2014",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2013-08-30",
      "reason": "The low density polyethylene (LDPE) bag used to package implants adheres to the highly polished surface of the devices.",
      "applies": "this device",
      "via": "K953337"
    },
    {
      "number": "Z-2290-2010",
      "classification": "",
      "status": "terminated",
      "initiated": "2010-07-05",
      "reason": "Zimmer, Inc. has determined that there is a potential for a package to be labeled as a 28 mm femoral\nhead but contain a 32 mm femoral head. In addition, there is a potential for the patient record label to\nbe incorrect.",
      "applies": "this device",
      "via": "K953337"
    },
    {
      "number": "Z-2291-2010",
      "classification": "",
      "status": "terminated",
      "initiated": "2010-07-05",
      "reason": "Zimmer, Inc. has determined that there is a potential for a package to be labeled as a 28 mm femoral\nhead but contain a 32 mm femoral head. In addition, there is a potential for the patient record label to\nbe incorrect.",
      "applies": "this device",
      "via": "K953337"
    },
    {
      "number": "Z-0198-2010",
      "classification": "",
      "status": "terminated",
      "initiated": "2009-08-03",
      "reason": "Manufacturing material may be present on the surface of the implant.",
      "applies": "this device",
      "via": "K953337"
    },
    {
      "number": "Z-1066-05",
      "classification": "",
      "status": "terminated",
      "initiated": "2005-04-15",
      "reason": "May be mislabeled as to dimensions; may be 12/14 taper, 28mm dia., neck length +3.5mm.",
      "applies": "this device",
      "via": "K953337"
    }
  ],
  "eu_market_availability": null,
  "properties": [
    {
      "group": "Identity",
      "fields": [
        {
          "label": "Generic name",
          "values": [
            {
              "value": "Metallic femoral head prosthesis",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "Model / Ref / Catalog",
          "values": [
            {
              "value": "9026-21",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            },
            {
              "value": "00-9026-021-00",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        }
      ]
    },
    {
      "group": "Identifiers",
      "fields": [
        {
          "label": "Primary DI / UDI-DI",
          "values": [
            {
              "value": "00889024155978",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "510(k) Number",
          "values": [
            {
              "value": "K840643",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            },
            {
              "value": "K953337",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "FDA product code",
          "values": [
            {
              "value": "LPH",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            },
            {
              "value": "JDI",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "GMDN term",
          "values": [
            {
              "value": "Metallic femoral head prosthesis",
              "code": "33175",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        }
      ]
    },
    {
      "group": "Classification",
      "fields": [
        {
          "label": "Device class",
          "values": [
            {
              "value": "Class II",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "Regulation number",
          "values": [
            {
              "value": "888.3358",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            },
            {
              "value": "888.3350",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "GUDID distribution status",
          "values": [
            {
              "value": "In Commercial Distribution",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        }
      ]
    },
    {
      "group": "Properties",
      "fields": [
        {
          "label": "Supplied sterile",
          "values": [
            {
              "value": "Yes",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "Single use",
          "values": [
            {
              "value": "Yes",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "Implantable",
          "values": [
            {
              "value": "Yes",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "Prescription use",
          "values": [
            {
              "value": "Yes",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "Lot / batch number",
          "values": [
            {
              "value": "Yes",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "Expiration date",
          "values": [
            {
              "value": "Yes",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "Sterilise before use",
          "values": [
            {
              "value": "No",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "Contains latex",
          "values": [
            {
              "value": "No",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "Over-the-counter",
          "values": [
            {
              "value": "No",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "Convenience kit",
          "values": [
            {
              "value": "No",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "Combination product",
          "values": [
            {
              "value": "No",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "HCT/P",
          "values": [
            {
              "value": "No",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "Premarket exempt",
          "values": [
            {
              "value": "No",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "Serial number",
          "values": [
            {
              "value": "No",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "Manufacturing date",
          "values": [
            {
              "value": "No",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        }
      ]
    }
  ]
}