Na 02.210.148
by Synthes
Listed
Valid until
Not applicable
FDA recalls
None ongoingDocuments
None
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | GUDID listing | Number10886982053176 | Date14 Sep 2015 | StatusListed |
Recalls
-
- Terminated Z-2389-2016
It was reported that the specified lots of the 2.4mm Titanium Locking Screw Self-Tapping with Stardrive Recess 12mm can break between the head and the shaft. This resulted from the neck feature being too close to the StarDrive feature leaving a thin wall between neck and drive.
-
- Terminated Z-0323-2016
It was discovered that the above part number and lot of 2.4mm Locking Screws may contain the incorrect package insert (GP2613, VA-LCP Distal Radius System). The correct package insert for the above part is GP2615 (Modular Mini Fragment LCP System).
-
- Terminated Z-0956-2013
It was discovered that one of the VA-LCP Dorsal Distal Radius Plates was incorrectly etched as Part Number 02.115.431 (110 degree) instead of the correct Part Number 02.115.231 (90 degree).
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — GUDID listing Listed
Specification
- Generic Name
- Orthopaedic bone screw (non-sliding)
- Model / Ref / Catalog
- 02.210.148, 02210148
- Primary DI / UDI-DI
- 10886982053176
- FDA Product Code
- HWC, HRS
- GMDN Term
- Orthopaedic bone screw (non-sliding)
- Device Class
- Class II
- Regulation Number
- 888.3040, 888.3030
- GUDID distribution status
- In Commercial Distribution
- Sterilization Method
- Moist Heat or Steam Sterilization
Other devices from Synthes
Properties
Other properties (10)
- Sterile: No
- Contains Latex: No
- Over-the-Counter: No
- Convenience Kit: No
- Combination Product: No
- HCT/P: No
- Premarket Exempt: No
- Serial Number: No
- Expiration Date: No
- Manufacturing Date: No