05.001.012
by Synthes GmbH
05.001.012 (05001012) is a surgical power tool hand-switch (FDA Class II) made by Synthes GmbH, cleared under FDA 510(k) K043310 on 5 Jan 2005 and listed in FDA GUDID since 30 Jun 2016 (UDI-DI 10886982128249), active as of 12 Nov 2026. 6 FDA recalls are linked to it.
Active
Valid until
No expiry
FDA recalls
None ongoingDocuments
Trust
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | GUDID listing | Number10886982128249 | Date30 Jun 2016 | StatusListed |
| United States | FDA 510(k)SESE | NumberK043310 | Date5 Jan 2005 | StatusActive |
Official records
- AccessGUDID record 10886982128249
- FDA 510(k) database K043310
From the FDA record
- Device description
- HAND SWITCH FOR ELECTRIC PEN DRIVE
- MRI safety
- Labeling does not contain MRI Safety Information
- Natural rubber latex
- No latex statement on the label
- Sterility
- Supplied non-sterile; sterilise before use (Moist Heat or Steam Sterilization)
Across registries
- FDA 510(k) K043310 decided 5 Jan 2005 Active
- FDA GUDID UDI-DI 10886982128249 listed 30 Jun 2016 Listed
Recalls
-
- Terminated Z-0681-2017
Package may be incorrectly labeled as containing 0.76 mm Drill Bit, but in fact contain a 1.1 mm diameter Drill Bit.
-
- Terminated Z-2784-2016
- Terminated Z-2783-2016
There was a potential for the Adaptor and Light Adaptor for Small Battery Drive and Small Battery Drive II to generate excessive internal pressure that may cause the products to burst.
-
- Terminated Z-2199-2016
DePuy Synthes is voluntarily initiating this recall due to affected devices that were distributed, etched and or labeled with MR Safety information not meeting the current ASTM standard.
-
- Terminated Z-1313-2015
In May 2013, the original recall was initiated due to complaints the Hand Switch for the Electric Pen Drive was not labeled correctly, and the on and lock labels were interchanged.
-
- Terminated Z-0615-2014
Synthes received a complaint from a customer in which the Hand Switch for the Electric Pen Drive was not labeled correctly. The "on" and "lock" labels were interchanged.
Timeline
- Cleared FDA 510(k) K043310
- Recall initiated FDA recalls Z-0615-2014
- Recall initiated FDA recalls Z-1313-2015
- Recall initiated FDA recalls Z-2199-2016
- Recall initiated FDA recalls Z-2784-2016
- Recall initiated FDA recalls Z-2783-2016
- Listed FDA GUDID 10886982128249
- Recall initiated FDA recalls Z-0681-2017
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- Generic name
- Surgical power tool hand-switch
- Model / Ref / Catalog
- 05.001.012, 05001012
- Primary DI / UDI-DI
- 10886982128249
- 510(k) Number
- FDA product code
- DZI, ERL, HWE, HBE
- GMDN term
- Surgical power tool hand-switch
- Device class
- Class II
- Regulation number
- 872.4120, 874.4250, 878.4820, 882.4310
- GUDID distribution status
- In Commercial Distribution
- Sterilisation method
- Moist Heat or Steam Sterilization