Na 170730
Listed
Valid until
Not applicable
FDA recalls
None ongoingDocuments
None
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | GUDID listing | Number08717213058118 | Date13 Nov 2025 | StatusListed |
Recalls
-
- Terminated Z-0987-2017
Cross profile for Varian 60 degree wedge shows "horns."
-
- Terminated Z-0604-2016
When using the option "Tumor Overlap Fraction" in VMAT planning it has been observed that in rare cases the system does include an organ at risk as target volume. This could result in open MLC, and open jaws in areas away from the target volume.
-
- Terminated Z-1712-2015
Incorrect dose calculation for Regions of Interest (ROIs) defined on a secondary image series.
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — GUDID listing Listed
Specification
- Generic Name
- Radiation therapy software
- Model / Reference
- 170730
- Primary DI / UDI-DI
- 08717213058118
- 510(k) Number
- FDA Product Code
- MUJ
- GMDN Term
- Radiation therapy software
- Device Class
- Class II
- Regulation Number
- 892.5050
- GUDID distribution status
- In Commercial Distribution
Other devices from Nucletron B.V.
Properties
Other properties (12)
- Sterile: No
- Single Use: No
- Implantable: No
- Contains Latex: No
- Over-the-Counter: No
- Convenience Kit: No
- Combination Product: No
- HCT/P: No
- Premarket Exempt: No
- Serial Number: No
- Expiration Date: No
- Manufacturing Date: No