Skip to content

2311

by St. Jude Medical, Inc.

Part of a family of 13 devices: see the family page

Ref 2311

2311 is a neural-tissue electrical stimulation lead adaptor (EU MDR Class III; FDA Class III; FDA product code LGW, Stimulator, Spinal-Cord, Totally Implanted for Pain Relief) made by St. Jude Medical, Inc., registered in EUDAMED on 25 Aug 2026 (Basic UDI-DI 5415067ACC000254) with CE/MDR certificate MDR 779805 valid until 11 Dec 2028, approved under FDA PMA P010032 on 21 Nov 2001 and listed in FDA GUDID since 12 Mar 2015 (UDI-DI 05415067016119), active as of 25 Oct 2026. 14 FDA recalls are linked to it.

Active European Union and United States. Also listed in GUDID (US).

CE MDR PMA

Valid until

11 Dec 2028

CE MDR certificate

FDA recalls

8 ongoing

This device is named in the recall

Documents

1 certificate

Certifier BSI Group The Netherlands B.V.

PMA P010032

FDA submission, from GUDID

Trust

Unverified

Source EUDAMED, GUDID, synced 25 Oct 2026.

Regulatory status

as of registry sync 25 Oct 2026

RegionAuthorizationNumberDateStatus
European UnionCE/MDR EUDAMED Number5415067ACC000254 Basic UDI-DI 05415067016119UDI-DIDate25 Aug 2026RegisteredStatusActive
United StatesGUDID listing GUDID Number05415067016119 UDI-DI Date12 Mar 2015ListedStatusListed GUDID: in commercial distribution
United StatesFDA PMAAPPR GUDID from GUDID Approved NumberP010032 Date21 Nov 2001DecidedStatusActive

Official records

From the FDA record

Device description
8-channel adapter, M, 10cm
MRI safety
Labeling does not contain MRI Safety Information
Natural rubber latex
No latex statement on the label
Sterility
Supplied sterile

From the GUDID record (AccessGUDID), as read on 25 Oct 2026, as the labeller entered it.

Across registries

3 registries list this device

  • EUDAMED Basic UDI-DI 5415067ACC000254 registered 25 Aug 2026 Active· +1 more in the status table
  • FDA PMA P010032 decided 21 Nov 2001 Active
  • FDA GUDID UDI-DI 05415067016119 listed 12 Mar 2015 Listed

Certificates

1 certificate · EUDAMED

EUDAMED certificates (1)
  • MDR 779805 Active Supplemented

    Technical documentation EU MDR

    Certifier
    BSI Group The Netherlands B.V. 2797
    Issued
    25 Jun 2026
    Expires
    11 Dec 2028
    Reason
    Other

Recalls

14 recalls · 8 ongoing · FDA

    • Ongoing Z-0458-2026 FDA Class II initiated 6 Oct 2025

    There is a potential issue affecting Eterna Spinal Cord Stimulation (SCS) Implantable Pulse Generators (IPGs), Model 32400 and Liberta RC Deep Brain Stimulation (DBS) Implantable Pulse Generators (IPGs), Model 62400 . These devicess may experience a loss of communication with the Clinician Programmer (CP) and/or the Patient Controller (PC).

  • 7 recalls, same reason

    • Ongoing Z-2505-2023 FDA Class I initiated 18 Jul 2023
    • Ongoing Z-2506-2023 FDA Class I initiated 18 Jul 2023
    • Ongoing Z-2507-2023 FDA Class I initiated 18 Jul 2023
    • Ongoing Z-2508-2023 FDA Class I initiated 18 Jul 2023
    • Ongoing Z-2509-2023 FDA Class I initiated 18 Jul 2023
    • Ongoing Z-2510-2023 FDA Class I initiated 18 Jul 2023
    • Ongoing Z-2511-2023 FDA Class I initiated 18 Jul 2023

    Firm has received complaints from patients who are unable to exit MRI (Magnetic Resonance Imaging) mode on their IPGs (Implantable Pulse Generators).

  • 2 recalls, same reason

    • Terminated Z-2691-2017 FDA Class II initiated 2 Jun 2017
    • Terminated Z-2692-2017 FDA Class II initiated 2 Jun 2017

    The firm discovered a risk of loss of Neuromodulation therapy due to exposure of the IPG to monopolar electrosurgical devices (ESU) during medical procedures.

    • Terminated Z-1785-2013 FDA Class II initiated 30 Nov 2012

    There is issues of warmth or heating at the implant site during charging for the Eon Mini implantable pulse generators (IPGs). In the July 2012 letter St. Jude Medical informed customers that they would be implementing design improvements to the charger to address possible increased energy dissipation when the charger is misaligned with the IPG or the IPG is implanted too near the surface of the

    • Terminated Z-0025-2013 FDA Class II initiated 4 Sep 2012

    Eon Mini IPGs manufactured in April 2012 could potentially exhibit a sudden, brief surge in stimulation that would be felt by the patient as uncomfortable or painful. SJMs routine testing, monitoring, and analysis of manufacturing processes identified three lots of Eon Mini IPGs that the internal battery has the potential to come into contact with the internal microcontroller board.

    • Terminated Z-1738-2013 FDA Class II initiated 26 Jul 2012

    St. Jude Medical has informed your doctor that a number of cases have been reported in which discomfort associated with heating occurred at the device site while patients were using the Charging System to charge their spinal cord stimulator. In three cases, patients received a burn to the skin (one second-degree and two first-degree burns) at the implant site. As of June 30, 2012, St. Jude Medica

    • Terminated Z-3091-2011 FDA Unclassified (pre-2012) initiated 2 Jun 2009

    Product programmed with unapproved software was distributed.

Timeline

19 dated events

  1. Recall initiated FDA recalls Z-2506-2023
  2. Recall initiated FDA recalls Z-2507-2023
  3. Recall initiated FDA recalls Z-2508-2023
  4. Recall initiated FDA recalls Z-2509-2023
  5. Recall initiated FDA recalls Z-2510-2023
  6. Recall initiated FDA recalls Z-2511-2023
  7. Recall initiated FDA recalls Z-0458-2026
  8. Certificate issued EUDAMED MDR 779805
  9. Registered EUDAMED 5415067ACC000254
  10. Certificate expires EUDAMED MDR 779805

9 earlier events are in the sections above.

Coverage

2 jurisdictions

United States of AmericaSwedenFinlandLatviaEstoniaDenmarkLithuaniaIrelandNetherlandsGermanyPolandFranceBelgiumLuxembourgCzech RepublicSlovakiaAustriaHungaryRomaniaSloveniaBulgariaPortugalSpainItalyCroatiaGreeceMaltaCyprus
  • Active
  • Expiring
  • Withdrawn
  • Listed
  • Not covered
  • European Union CE/MDR Active
  • United States FDA PMA Active

Specification

EUDAMED · GUDID

Generic name
Neural-tissue electrical stimulation lead adaptor
Device name
Neuromodulation Accessories
Model / Ref / Catalog
2311
Primary DI / UDI-DI
05415067016119
Basic UDI-DI
5415067ACC000254
PMA number
FDA product code
LGW
EMDN code
J0280 Implantable neurostimulators - accessories
GMDN term
Neural-tissue electrical stimulation lead adaptor
Device class
Class III
EU registration type
Standard device
EU market of registration
Belgium
GUDID distribution status
In Commercial Distribution
Size
Length — 10 Centimeter

Similar devices in this category

Other devices filed under IMPLANTABLE NEUROSTIMULATORS.

Other devices from St. Jude Medical, Inc.

See all devices from St. Jude Medical, Inc. →

Properties

  • Supplied sterile
  • Single use
  • Implantable
  • Contains CMR substances (carcinogenic, mutagenic or toxic to reproduction)
  • Prescription use
  • Lot / batch number
  • Serial number
  • Expiration date
Other properties (17, all answered No)