Na 294.771
by Synthes United States GUDID
Product code LXT
Regulatory status
Recall ongoing. 1 ongoing FDA recall. See recalls
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | GUDID listing | Number10886982181824 | Date14 Sep 2015 | StatusIn distribution |
Coverage
- Listed
- Not covered
United States — GUDID listing Listed
Recalls
-
- Ongoing Z-1816-2013
It was discovered internally that there was a discrepancy between the insert provided with the Tube to Tube Clamp and what was called for in the Bill of Materials. Following a detailed investigation, it was determined that the incorrect insert was used which indicates that the device is MR Safe, when it is actually MR Conditional.
-
- Terminated Z-0043-2018
The item is being recalled due to an inadequate pouch seal. This defect has the potential to lead to a loss of sterile pouch integrity, which could impact the products sterility assurance and may lead to serious adverse events such as infections, both localized and systemic.
-
- Terminated Z-2451-2015
Labeling changes were made related to MR conditions that align with standard definitions for MR Safe, MR Conditional, and MR Unsafe. This recall identifies additional Trauma External Fixation System part numbers that should be considered MR Conditional. One serious injury reported.
-
- Terminated Z-1833-2014
Labeling changes have been made related to MR (Magnetic Resonance imaging) conditions as a result of changes in required protocols to designate a product MR Safe, MR Conditional, or MR Unsafe.
-
- Terminated Z-0941-2013
The L1 hex nuts, used to tighten the Large Combination Clamp, were oversized and the 11 mm sockets and wrenches will not fit on the hex nuts.
-
- Terminated Z-0277-2013
The insert provided with the product for the specific lot number indicates that this product is MR safe. This product is MR Conditional and is etched as MR Conditional. The insert is incorrect.
Specification
- Trade Name
- Na
- Generic Name
- Orthopaedic bone screw (non-sliding)
- Model / Ref / Catalog
- 294.771, 294771
- Primary DI / UDI-DI
- 10886982181824
- FDA Product Code
- LXT, KTT, HWC
- GMDN Term
- Orthopaedic bone screw (non-sliding)
- Device Class
- Class II
- Regulation Number
- 888.3030, 888.3040
- Commercial Distribution
- In Commercial Distribution
- Sterilization Method
- Moist Heat or Steam Sterilization
- MRI Safety
- MR Conditional
Other devices from Synthes
Properties & raw data
Show properties, registry sources and raw records
Other properties (10)
- Sterile: No
- Contains Latex: No
- Over-the-Counter: No
- Convenience Kit: No
- Combination Product: No
- HCT/P: No
- Premarket Exempt: No
- Serial Number: No
- Expiration Date: No
- Manufacturing Date: No
Sources (1)
- FDA UDI 7c21abda-a1b6-442b-8c53-bb675d051637