47-2480-026-41 is a medical device (FDA Class II; FDA product code JDI, Prosthesis, hip, semi-constrained, metal/polymer, cemented) made by Zimmer, Inc., cleared under FDA 510(k) K112885 on 19 Dec 2011 and listed in FDA GUDID since 24 Sep 2015 (UDI-DI 00889024093034), active as of 23 Oct 2026.

Active United States. Also listed in GUDID (US).

510(k)

Valid until

No expiry

510(k) clearance is not time-limited

FDA recalls

None ongoing

15 closed

Documents

510(k) K112885

FDA submission, from GUDID

Trust

Unverified

Source GUDID, synced 23 Oct 2026.

Regulatory status

as of registry sync 23 Oct 2026

RegionAuthorizationNumberDateStatus
United StatesGUDID listing Number00889024093034 UDI-DI Date24 Sep 2015ListedStatusListed GUDID: in commercial distribution
United StatesFDA 510(k)SESE from GUDID Substantially Equivalent · Traditional NumberK112885 Date19 Dec 2011DecidedStatusActive

Official records

From the FDA record

MRI safety
Labeling does not contain MRI Safety Information
Natural rubber latex
No latex statement on the label
Sterility
Supplied sterile

From the GUDID record (AccessGUDID), as read on 23 Oct 2026, as the labeller entered it.

Across registries

2 registries list this device

  • FDA 510(k) K112885 decided 19 Dec 2011 Active
  • FDA GUDID UDI-DI 00889024093034 listed 24 Sep 2015 Listed

Recalls

15 recalls · FDA

    • Terminated Z-1677-2016 FDA Class II initiated 4 May 2016 on shared 510(k) K112885

    A small hole may be present in the inner tray of specific lots of sterile Hip and Trauma Implant Screws.

  • 4 recalls, same reason

    • Terminated Z-1102-2016 FDA Class II initiated 11 Jan 2016 on shared 510(k) K112885
    • Terminated Z-1108-2016 FDA Class II initiated 11 Jan 2016 on shared 510(k) K112885
    • Terminated Z-1111-2016 FDA Class II initiated 11 Jan 2016 on shared 510(k) K112885
    • Terminated Z-1116-2016 FDA Class II initiated 11 Jan 2016 on shared 510(k) K112885

    As a result of the insufficient sealer calibration data for product packaged in firm's Building II between August 2010 and April 2013. The affected products are sterile.

  • 8 recalls, same reason

    • Terminated Z-2244-2014 FDA Class II initiated 22 Jul 2014 on shared 510(k) K112885
    • Terminated Z-2245-2014 FDA Class II initiated 22 Jul 2014 on shared 510(k) K112885
    • Terminated Z-2246-2014 FDA Class II initiated 22 Jul 2014 on shared 510(k) K112885
    • Terminated Z-2247-2014 FDA Class II initiated 22 Jul 2014 on shared 510(k) K112885
    • Terminated Z-2248-2014 FDA Class II initiated 22 Jul 2014 on shared 510(k) K112885
    • Terminated Z-2249-2014 FDA Class II initiated 22 Jul 2014 on shared 510(k) K112885
    • Terminated Z-2250-2014 FDA Class II initiated 22 Jul 2014 on shared 510(k) K112885
    • Terminated Z-2251-2014 FDA Class II initiated 22 Jul 2014 on shared 510(k) K112885

    Zimmer is recalling various trauma screws that failed a leak test due to damage of the inner cavity only.

  • 2 recalls, same reason

    • Terminated Z-1040-2014 FDA Class II initiated 11 Oct 2013 on shared 510(k) K112885
    • Terminated Z-1042-2014 FDA Class II initiated 11 Oct 2013 on shared 510(k) K112885

    Zimmer recently conducted a review of all historic packaging validations at its Zimmer Manufacturing B.V. (ZMBV) facility in Ponce, Puerto Rico. Based on this review, it was determined that the packaging operations conducted in the Building 1 packaging operation at this facility were not properly validated. As a result, Zimmer conducted seal integrity testing to evaluate if previously packaged product from this facility met the established requirements for sterile barrier integrity. The acceptance criteria established for this testing was 99% reliability with a 90% confidence interval. These packaging tray configurations met the acceptance criteria for sterile barrier integrity of both the inner and outer trays, however, Zimmer has decided to recall all manufacturing lots that have not expired and that were sterile packaged in the ZMBV facility Building 1 packaging operations.

Recalls on a shared submission name another device cleared under the same FDA submission as this one.

Timeline

17 dated events

  1. Recall initiated FDA recalls Z-2248-2014
  2. Recall initiated FDA recalls Z-2249-2014
  3. Recall initiated FDA recalls Z-2250-2014
  4. Recall initiated FDA recalls Z-2251-2014
  5. Listed FDA GUDID 00889024093034
  6. Recall initiated FDA recalls Z-1102-2016
  7. Recall initiated FDA recalls Z-1108-2016
  8. Recall initiated FDA recalls Z-1111-2016
  9. Recall initiated FDA recalls Z-1116-2016
  10. Recall initiated FDA recalls Z-1677-2016

7 earlier events are in the sections above.

Coverage

1 jurisdiction

United States of America
  • Active
  • Expiring
  • Withdrawn
  • Listed
  • Not covered

United States — FDA 510(k) Active

Specification

GUDID

Generic name
Orthopaedic bone screw (non-sliding)
Model / Ref / Catalog
47-2480-026-41
Primary DI / UDI-DI
00889024093034
510(k) Number
FDA product code
JDI
GMDN term
Orthopaedic bone screw (non-sliding)
Device class
Class II
Regulation number
888.3350
GUDID distribution status
In Commercial Distribution
Size
Length — 26 Millimeter

Similar devices in this category

Other devices filed under Orthopaedic bone screw (non-sliding).

Other devices from Zimmer, Inc.

See all devices from Zimmer, Inc. →

Properties

  • Supplied sterile
  • Single use
  • Implantable
  • Prescription use
  • Lot / batch number
  • Expiration date
Other properties (9, all answered No)