Na 5265300-20
by GE Medical Systems, L.L.C. United States GUDID
Product code LNH
Regulatory status
Recall ongoing. 16 ongoing FDA recalls. See recalls
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | GUDID listing | Number00840682102728 | Date1 Aug 2016 | StatusIn distribution |
Coverage
- Listed
- Not covered
United States — GUDID listing Listed
Recalls
-
- Ongoing Z-2306-2025
- Ongoing Z-2301-2025
- Ongoing Z-2302-2025
GE HealthCare has become aware that for certain MR systems (SIGNA Artist, SIGNA Voyager, Discovery MR450 1.5T, Optima MR450w 1.5T, SIGNA Architect, SIGNA Premier, SIGNA UHP, Discovery MR750 3.0T, Discovery MR750w 3.0T, SIGNA Hero, SIGNA PET/MR, SIGNA Artist Evo, SIGNA Voyager AIR (China Only), SIGNA Architect AIR (China Only), SIGNA Premier XT (China Only), SIGNA Premier Elite (China Only), SIGNA Premier MAX (China Only), SIGNA Hero XT (China Only), SIGNA Hero Elite (China Only), SIGNA Hero MAX (China Only), SIGNA PET/MR AIR (China Only), SIGNA PET/MR AIR Plus (China Only), the planned maintenance steps to check the functionality of the caster locks on the MR patient table may not have been performed.
-
- Ongoing Z-0517-2025
Gradient coils for certain MR systems (see affected product list in this letter), under specific conditions, can produce elevated acoustic noise during scanning.
-
- Ongoing Z-2464-2024
- Ongoing Z-2468-2024
- Ongoing Z-2467-2024
- Ongoing Z-2465-2024
GE HealthCare has become aware that for certain MR systems, the system predicted B1+RMS value can exceed the B1+RMS user prescribed limit when scanning in Low SAR Mode and the following two conditions are met: 1. 2D FSE T2 FLAIR OR T2 FLAIR Propeller imaging sequence is selected, AND 2. the Optimized T2 FLAIR sequence option is turned off OR not included in the MR configuration. If this occurs, it can result in overheating of an MR conditional implant.
-
- Ongoing Z-0593-2022
- Ongoing Z-0602-2022
- Ongoing Z-0594-2022
- Ongoing Z-0601-2022
The magnetic resonance systems could potentially have a cryogen ventilation system that does not meet the venting requirements.
-
- Ongoing Z-2055-2021
- Ongoing Z-2064-2021
- Ongoing Z-2063-2021
- Ongoing Z-2056-2021
GE Healthcare has recently become aware of an issue on the affected products listed below where the "Patient Orientation" button may inadvertently be clicked when intending to click on the "Save RX" button. This changes the prescribed patient orientation on the system prior to running the initial 3-Plane Localizer Scan (See Figure 1 for reference to buttons.) Selecting and saving a patient orientation that does not match the patient's actual position may result in incorrectly annotated and/or flipped images.
-
- Terminated Z-0135-2022
- Terminated Z-0136-2022
- Terminated Z-0140-2022
Under certain conditions, some slices may be missing which can lead to a gap of anatomy in the 3D Volume images.
-
- Terminated Z-3279-2018
GE Medical Systems has discovered a small area on the bore surface of the Discovery MR750w Nuclear Magnetic Resonance Imaging System, normally accessible to the patient, can get warm enough to cause a serious burn in the event the patient touches the bore and proper padding is not used.
-
- Terminated Z-3025-2017
- Terminated Z-3023-2017
- Terminated Z-3022-2017
- Terminated Z-3024-2017
Possible incorrect wiring of the Uninterruptable Power Supply, which could contribute to a potential shock hazard.
-
- Terminated Z-1188-2017
Potential safety issue with the patient bore heating on the Discovery MR450 narrow bore 1.5T MRI. In the event of a RF component failure, typically happening after extended system use, a small area on the bore surface may become warmer than normal. If this happens in an area where there is patient contact with the bore, and proper padding is not used, it may cause a serious burn.
-
- Terminated Z-2131-2016
GE Healthcare has recently become aware of a potential safety issue with the patient bore heating on the Discovery MR750w. A small area on the bore surface, which is normally accessible to the patient, can get warm enough to cause a serious burn in the event the patient touches the bore and proper padding is not used. No injury has been reported to GE as a result of this issue.
-
- Terminated Z-0201-2019
GE Healthcare has recently become aware of a potential safety issue with the 3.0T 6 Channel Flex Coil used with the 3T MR750W Surgical Suite Scanners. Coil overheating can occur when the device is used in Mode 2 setup. This could lead to a serious patient thermal injury. There have been no injuries reported as a result of this issue. UPDATE 5/27/2016: Recall has been updated to include an additional product which includes the Discovery 3.0T MR750 systems running Application Software DV24 R01, DV25 R02, or DV25.1 M3.
Specification
- Trade Name
- Na
- Generic Name
- Full-body MRI system, superconducting magnet
- Model / Reference
- 5265300-20
- Primary DI / UDI-DI
- 00840682102728
- 510(k) Number
- FDA Product Code
- LNH
- GMDN Term
- Full-body MRI system, superconducting magnet
- Device Class
- Class II
- Regulation Number
- 892.1000
- Commercial Distribution
- In Commercial Distribution
Other devices from GE Medical Systems, L.L.C.
Properties & raw data
Show properties, registry sources and raw records
Other properties (11)
- Sterile: No
- Single Use: No
- Implantable: No
- Contains Latex: No
- Over-the-Counter: No
- Convenience Kit: No
- Combination Product: No
- HCT/P: No
- Premarket Exempt: No
- Lot / Batch Number: No
- Expiration Date: No
Sources (1)
- FDA UDI a5205581-fc48-4c09-8a14-723901524298