Na 6504
by Boston Scientific United States GUDID
Product code LWS
Regulatory status
Recall ongoing. 3 ongoing FDA recalls. See recalls
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | GUDID listing | Number00802526489235 | Date24 Sep 2014 | StatusIn distribution |
Coverage
- Listed
- Not covered
United States — GUDID listing Listed
Recalls
-
- Ongoing Z-0673-2024
The software application on the Programmer/Recorder/Monitor (PRM) for Implantable Pulse Generators will display certain diagnostic dates incorrectly, with a year in the 1990s.
-
- Ongoing Z-1216-2023
Under specific circumstances, the U.S. product registration system did not send up enablement requests to U.S. LATITUDE Customer Support during an 8-month interval in 2022. For these patients, the HeartLogic Index and any associated Yellow Alerts are currently not visible in LATITUDE as may have been intended by the healthcare provider..
-
- Ongoing Z-1669-2022
There is potential for unanticipated software anomalies to occur when using the Model 2868 Application loaded on the 3120 ZOOM LATITUTDE programmer when communicating with certain Boston Scientific transvenous defibrillators.
-
- Terminated Z-3260-2018
There is a potential for the Pacing System Analyzer (PSA) to exhibit unintended cross-chamber stimulation.
-
- Terminated Z-1445-2012
Boston Scientific LATITUDE Patient Management System Communicators were shipped to the incorrect patients.
-
- Terminated Z-1096-2012
Boston Scientific is conducting a recall involving the LATITUDE Patient Management System - Model 6488, software version 7.2 because patients are not able to perform setup. There is no risk to health associated with this issue.
-
- Terminated Z-0333-2012
Current Boston Scientific defibrillator systems measure shock lead impedances up to 200 ohms, and export results to the programmer and LATITUDE monitoring system via the Daily Measurements feature. A measured shock lead impedance that is greater than 125 ohms or less than 20 ohms will generate an alert message on programmer screens and a Red Alert within LATITUDE (if activated). However, the mea
-
- Terminated Z-0037-2012
Boston Scientific has determined that under certain conditions, LATITUDE Paceart Integration (LPI) software may not properly transfer data resulting in the incorrect summary of programmed parameter values in the ICD Summary tab display and Current Programming Tachy summary section of the Paceart report.
-
- Terminated Z-1631-2010
Boston Scientific has determined that alert conditions for a limited subset of out-of-range Daily Measurements related to leads may not be recognized by Model 6476 (United States) or Model 6468 (Europe) LATITUDE in-home communicators. If a communicator fails to recognize a valid alert situation, notification for this or subsequent alerts for the same test will not be displayed for physician review
-
- Terminated Z-0002-2007
Final software load did not occur prior to shipment of select programmers.
Specification
- Trade Name
- Na
- Generic Name
- Pacing/defibrillation lead cap kit
- Model / Reference
- 6504
- Primary DI / UDI-DI
- 00802526489235
- PMA Number
- FDA Product Code
- LWS
- GMDN Term
- Pacing/defibrillation lead cap kit
- Device Class
- Class III
- Commercial Distribution
- In Commercial Distribution
Other devices from Boston Scientific
Properties & raw data
Show properties, registry sources and raw records
Other properties (9)
- Implantable: No
- Contains Latex: No
- Over-the-Counter: No
- Convenience Kit: No
- Combination Product: No
- HCT/P: No
- Premarket Exempt: No
- Serial Number: No
- Manufacturing Date: No
Sources (1)
- FDA UDI ace1d544-69f6-47f2-b8d7-0413023c6a82