71932826
71932826 is a device of GMDN type press-fit femoral stem prosthesis (FDA Class II; FDA product code JDI, Prosthesis, hip, semi-constrained, metal/polymer, cemented) made by Smith & Nephew, Inc., cleared under FDA 510(k) K240783 on 21 Oct 2024 (2 earlier submissions) and listed in FDA GUDID since 29 Aug 2015 (UDI-DI 00885556060193), active as of 23 Oct 2026. One FDA recall is linked to it.
Active
Valid until
No expiry
FDA recalls
1 ongoingDocuments
Trust
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK240783 | Date21 Oct 2024 | StatusActive |
| United States | GUDID listing | Number00885556060193 | Date29 Aug 2015 | StatusListed |
Official records
- AccessGUDID record 00885556060193
- FDA 510(k) database K240783
- FDA 510(k) database K211176
- FDA 510(k) database K052792
From the FDA record
- Device description
- SHORT TAPER STEM PC STD SIZE 10
- MRI safety
- Labeling does not contain MRI Safety Information
- Natural rubber latex
- No latex statement on the label
- Sterility
- Supplied sterile
Across registries
- FDA 510(k) K240783 decided 21 Oct 2024 Active
- FDA GUDID UDI-DI 00885556060193 listed 29 Aug 2015 Listed
Recalls
-
- Ongoing Z-0326-2025
The Bipolar assembly contains an oversized Retainer Ring.
-
- Terminated Z-1921-2018
Three lots of Anthology HO Porous size 7 stems were manufactured with an undersized taper length. The device taper length was incorrectly manufactured as 0.393mm opposed to 0.442 mm.
Timeline
- Cleared FDA 510(k) K052792
- Listed FDA GUDID 00885556060193
- Recall initiated FDA recalls Z-1921-2018
- Cleared FDA 510(k) K211176
- Recall initiated FDA recalls Z-0326-2025
- Cleared FDA 510(k) K240783
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- Generic name
- Press-fit femoral stem prosthesis
- Model / Ref / Catalog
- 71932826
- Primary DI / UDI-DI
- 00885556060193
- FDA product code
- JDI
- GMDN term
- Press-fit femoral stem prosthesis
- Device class
- Class II
- Regulation number
- 888.3350
- GUDID distribution status
- In Commercial Distribution
Similar devices
Other devices filed under FDA product code JDI (Prosthesis, hip, semi-constrained, metal/polymer, cemented).
- MP Reconstruction Prosthesis 172-950/20 by WALDEMAR LINK GmbH & Co. KG
- POROCOAT DUAL-LOCK TOTAL HIP SYSTEM by Depuy, Inc.
- CoCr Femoral Ball Head 01.25.011 by Medacta International SA
- G7® Acetabular System 110017403 by Biomet Orthopedics, Inc.
- Continuum® Trilogy® Allofit® 00-8731-016-36 by Zimmer, Inc.
- G7® 010002485 by Biomet Orthopedics, Inc.
- Pantheon System Salvage Stem (Uncemented) 0200515 by ADLER ORTHO SPA
- Continuum® Trilogy® Allofit® 00-8734-005-28 by Zimmer, Inc.
- MODULAR HIP SYSTEM by Smith & Nephew Richards, Inc.
- Metafix Hip Stem 3997045 by CORIN LTD