Na 8596SC
by Medtronic, Inc. United States GUDID
Product code LKK
Regulatory status
Recall ongoing. 6 ongoing FDA recalls. See recalls
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | GUDID listing | Number00763000916633 | Date6 May 2024 | StatusIn distribution |
Coverage
- Listed
- Not covered
United States — GUDID listing Listed
Recalls
-
- Ongoing Z-2201-2026
Software issue: When programming the SynchroMed Pump with the A810 CP App in Flex Infusion Mode, the infusion steps should automatically be arranged in consecutive order, from earliest to latest. However, in rare cases, the steps may be displayed out of order. If this occurs, the programmed Flex Infusion schedule will not be delivered at the intended time, if at all.
-
- Ongoing Z-1506-2025
Product complaints were received describing the A820 myPTM app taking longer than expected for patients to interact with their implantable pump.
-
- Ongoing Z-2352-2024
Design update to the Ascenda Intrathecal Catheter Models 8780, 8781, and 8784. The intent of the design update is to reduce the potential for tissue growth into the Ascenda catheter connector which may potentially lead to catheter occlusion.
-
- Ongoing Z-0615-2024
- Ongoing Z-0614-2024
Medtronic recently identified that if the SynchroMed II pump switches into telemetry mode due to electromagnetic interference (EMI) from an MRI scan, while the pump is sounding an alarm, the pump will not resume drug delivery after leaving the MRI magnetic field, which is inconsistent with the current labeling.
-
- Ongoing Z-0896-2018
Medtronic implemented a final design change intended to address the motor stall due to corrosion and is recalling devices with the previous design that are not implanted.
-
- Completed Z-0522-2024
The previous software application version (1.1.300) is missing a decimal separator (a comma) for parameter range guidance values displayed on some of the programming screens: Catheter, Reservoir, Infusion, Bolus, myPTM, and Alarm.
-
- Terminated Z-0508-2020
There is a potential for the presence of a foreign particle inside the pump motor assembly which could interfere with the motor gear rotation and lead to a permanent motor stall.
-
- Terminated Z-0537-2018
There is a possibility that specific lots of catheters that have recently been distributed may be at risk for an increased potential for kinking at the proximal end where the catheter connects to the drug infusion pump.
-
- Terminated Z-0788-2017
Medtronic is following up to a May 2013 communication regarding the Priming Bolus function for the SynchroMed Infusion System. Medtronic is updating the Model 8870 software application card (to version AAU01) and the SynchroMed pump labeling to address the priming bolus issue.
-
- Terminated Z-1271-2016
Medtronic Neuromodulation is initiating a voluntary removal of a single Model 8781 Ascenda" Intrathecal Catheter kit. We have identified that this kit may have been produced with incorrect outer package labels. As a result, it may not contain all components necessary to complete the implant procedure.
-
- Terminated Z-1681-2015
Medtronic is conducting a recall of a specific subset of Model 8637-20 and 8637-40 SynchroMed II implantable drug pumps because the audible alarm could cease to function.
-
- Terminated Z-1061-2011
- Terminated Z-1062-2011
- Terminated Z-1060-2011
This letter provides important reminders concerning the potential for a pocket fill during a SynchroMed II or SynchroMed EL implantable drug pump refill procedure, and important patient management recommendations that will be added to our product labeling. A pocket fill is the inadvertent injection of all or some of the prescribed drug into the patients subcutaneous tissue, which includes the pum
-
- Terminated Z-2276-2009
This recall is related to the potential for reduced battery performance in a small percentage of Medtronic Model 8637 SynchroMed II pumps with batteries manufactured during two distinct time periods prior to April 2005. Nature of the Device Issue: As part of ongoing analysis of returned explanted product, Medtronic has confirmed that reduced battery performance resulted in eight (8) occurrence
-
- Terminated Z-1214-2010
Routine monitoring of Medtronic's vigilance program identified user related programming errors. Medtronic has received reports of both overdose and underdose that were the result of programming use of SynchroMed implantable infusion pumps. These errors were most commonly associated with: --Priming Bolus Programming; and --Bridge Bolus Programming. Pump programming errors that result in overdos
-
- Terminated Z-2172-2008
- Terminated Z-2174-2008
- Terminated Z-2171-2008
- Terminated Z-2173-2008
The SC catheters cannot completely engage with the Model 700-04M portal connector. The result of this condition is a possible leakage or disconnect of the SC catheter from the portal during the continuous therapy trial period.
-
- Terminated Z-1147-2008
- Terminated Z-1144-2008
- Terminated Z-1154-2008
- Terminated Z-1149-2008
- Terminated Z-1143-2008
- Terminated Z-1151-2008
- Terminated Z-1153-2008
- Terminated Z-1150-2008
- Terminated Z-1148-2008
- Terminated Z-1152-2008
- Terminated Z-1145-2008
- Terminated Z-1142-2008
- Terminated Z-1146-2008
Device/Drug Interaction - The company updated the labeling for the devices to include current patient management and treatment recommendations. The company received reports of inflammatory mass formations at or near the distal tip of intrathecal catheters which infuse opioids, baclofen, or chemotherapy drugs into patients. On January 16, 2008, Medtronic sent a letter to doctors who implant these d
-
- Terminated Z-0739-2008
Pump motor stall due to gear shaft wear
-
- Terminated Z-1415-06
- Terminated Z-1414-06
Tip dislodgement during implantation-Medtronic is recalling Model 8731 Intrathecal Catheter and Model 8598 Intrathecal Catheter Distal Revision Kit because the platinum-iridium tip may be dislodged by the guide wire during implantation.
-
- Terminated Z-0391-05
Outer package of Lioresal Intrethecal Refill Kit, model 8561, identifies the kit to contain 1-20 ml ampule of baclofen injection for intrathecal administration, 500 mcg/ml (10 mg/20 ml). Lot N0016487 incorrectly contains 1-20 ml ampule of baclofen concentration 2000 mcg/ml (40 mg/20 ml).
-
- Terminated Z-1334-04
With the Model 8870 software application card versions AAA 02, BBB 04, BBC 02, and BBD 01, which is used in conjunction with the Model 8840 N'Vision Clinician Programmer, users may have mistakenly entered a periodic bolus interval into the minutes field, rather than the hours field, resulting in drug overdoses. This issue is limited to programming the SynchroMed and SynchroMed EL pumps.
-
- Terminated Z-1384-04
The sterile tray label for the Model 8540 Catheter Access Port kit may be incorrectly labeled with a Model 8551 Refill kit label. The outer box of the affected Model 8540 Catheter Access Port kit is labeled correctly and the component parts within the sterile tray are the correct components for a Model 8540 Catheter Access Port kit. The issue is limited to one lot of product: CS0293.
Specification
- Trade Name
- Na
- Generic Name
- Lumbar cerebrospinal fluid drainage catheterization kit
- Model / Reference
- 8596SC
- Primary DI / UDI-DI
- 00763000916633
- PMA Number
- FDA Product Code
- LKK
- GMDN Term
- Lumbar cerebrospinal fluid drainage catheterization kit
- Device Class
- Class III
- Commercial Distribution
- In Commercial Distribution
Other devices from Medtronic, Inc.
Properties & raw data
Show properties, registry sources and raw records
Other properties (9)
- Implantable: No
- Contains Latex: No
- Over-the-Counter: No
- Convenience Kit: No
- HCT/P: No
- Premarket Exempt: No
- Lot / Batch Number: No
- Expiration Date: No
- Manufacturing Date: No
Sources (1)
- FDA UDI 72fb64dd-456a-4b40-ba88-b37b14f3e1bc