Na ACU55790
by Sterilmed, Inc. GUDID
Listed (no authorization)
Regulatory status
Recall ongoing. 2 ongoing FDA recalls. See recalls
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | GUDID listing | Number10888551004749 | Date22 Sep 2016 | StatusIn distribution |
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not known
- Not covered
United States — GUDID listing Listed
Recalls
-
- Ongoing Z-0574-2026
- Ongoing Z-0573-2026
Affected catheters were reprocessed beyond validated cycles. This may compromise device sterility or mechanical integrity of the device and can lead to severe adverse events such as infection, vascular injury or embolism.
-
- Terminated Z-0551-2009
SteriMed, Inc is conducting a product removal of a small group of devices individually packaged and sealed within a pouch. The removal is being conducted because the possibility exists that some of the pouches were not properly sealed on one end. A breach in packaging seal or a failure in packaging integrity has the possibility of risk to the patient in terms of transmitting organisms capable
Specification
- Trade Name
- Na
- Generic Name
- Intracardiac ultrasound imaging catheter, steerable, reprocessed
- Model / Ref / Catalog
- ACU55790
- Primary DI / UDI-DI
- 10888551004749
- 510(k) Number
- FDA Product Code
- OWQ
- GMDN Term
- Intracardiac ultrasound imaging catheter, steerable, reprocessed
- Device Class
- Class II
- Regulation Number
- 870.1200
- Commercial Distribution
- In Commercial Distribution
- Sterilization Method
- Ethylene Oxide
Other devices from Sterilmed, Inc.
Properties
Other properties (9)
- Implantable: No
- Contains Latex: No
- Over-the-Counter: No
- Convenience Kit: No
- Combination Product: No
- HCT/P: No
- Premarket Exempt: No
- Serial Number: No
- Manufacturing Date: No