Na Express Coil - Posterior Array
by Ge Healthcare(Tianjin) Company Limited China GUDID
Product code LNI
Regulatory status
Recall ongoing. 3 ongoing FDA recalls. See recalls
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | GUDID listing | Number00840682118583 | Date1 Aug 2016 | StatusIn distribution |
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not known
- Not covered
United States — GUDID listing Listed
Recalls
-
- Ongoing Z-2477-2024
- Ongoing Z-2478-2024
GE HealthCare has become aware that for certain MR systems, the system predicted B1+RMS value can exceed the B1+RMS user prescribed limit when scanning in Low SAR Mode and the following two conditions are met: 1. 2D FSE T2 FLAIR OR T2 FLAIR Propeller imaging sequence is selected, AND 2. the Optimized T2 FLAIR sequence option is turned off OR not included in the MR configuration. If this occurs, it can result in overheating of an MR conditional implant.
-
- Ongoing Z-0610-2022
The magnetic resonance systems could potentially have a cryogen ventilation system that does not meet the venting requirements.
-
- Terminated Z-3021-2017
Possible incorrect wiring of the Uninterruptable Power Supply, which could contribute to a potential shock hazard.
-
- Terminated Z-1209-2013
GE Healthcare has recently become aware of a Reformat Image Flip Issue associated with GE MR Products. The axial and coronal reformatted images from a sagittal 3D series (such as with applications of FSPGR-3D T1, BRAVO T1, or CUBE T1) of over 126 slices may display a left-to-right image flip . In addition, the acquired sagittal images may be annotated with left-right location annotation revers
-
- Terminated Z-1433-2011
GE Healthcare has become aware of a potential issue associated with the gradient cables being inadvertently swapped while servicing the gradients on some GE Healthcare MR Systems that may impact patient safety. If the gradient cables are inadvertently swapped while servicing the gradients, and the Service Engineer does not perform the geometry check as required in the service procedure, images m
Specification
- Trade Name
- Na
- Generic Name
- MRI system coil, radio-frequency
- Model / Reference
- EXPRESS COIL - POSTERIOR ARRAY
- Primary DI / UDI-DI
- 00840682118583
- 510(k) Number
- FDA Product Code
- LNI, LNH
- GMDN Term
- MRI system coil, radio-frequency
- Device Class
- Class II
- Regulation Number
- 892.1000
- Commercial Distribution
- In Commercial Distribution
Other devices from Ge Healthcare(Tianjin) Company Limited
See all devices from Ge Healthcare(Tianjin) Company Limited →
Properties & raw data
Show properties, registry sources and raw records
Other properties (11)
- Sterile: No
- Single Use: No
- Implantable: No
- Contains Latex: No
- Over-the-Counter: No
- Convenience Kit: No
- Combination Product: No
- HCT/P: No
- Premarket Exempt: No
- Lot / Batch Number: No
- Expiration Date: No
Sources (1)
- FDA UDI 7780665f-d8cd-4351-86e2-67c7e126853f