Na G3341AA
by GE Medical Systems, L.L.C. United States GUDID
Product code LNH
Regulatory status
Recall ongoing. 5 ongoing FDA recalls. See recalls
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | GUDID listing | Number00840682115964 | Date1 Aug 2016 | StatusIn distribution |
Coverage
- Listed
- Not covered
United States — GUDID listing Listed
Recalls
-
- Ongoing Z-2463-2024
- Ongoing Z-2471-2024
GE HealthCare has become aware that for certain MR systems, the system predicted B1+RMS value can exceed the B1+RMS user prescribed limit when scanning in Low SAR Mode and the following two conditions are met: 1. 2D FSE T2 FLAIR OR T2 FLAIR Propeller imaging sequence is selected, AND 2. the Optimized T2 FLAIR sequence option is turned off OR not included in the MR configuration. If this occurs, it can result in overheating of an MR conditional implant.
-
- Ongoing Z-0860-2022
There is potential for the images to be flipped left to right.
-
- Ongoing Z-0603-2022
The magnetic resonance systems could potentially have a cryogen ventilation system that does not meet the venting requirements.
-
- Ongoing Z-2065-2021
GE Healthcare has recently become aware of an issue on the affected products listed below where the "Patient Orientation" button may inadvertently be clicked when intending to click on the "Save RX" button. This changes the prescribed patient orientation on the system prior to running the initial 3-Plane Localizer Scan (See Figure 1 for reference to buttons.) Selecting and saving a patient orientation that does not match the patient's actual position may result in incorrectly annotated and/or flipped images.
-
- Terminated Z-0141-2022
Under certain conditions, some slices may be missing which can lead to a gap of anatomy in the 3D Volume images.
-
- Terminated Z-0877-2020
It was identified that due to a potential installation workflow issue, the MR system date could be set incorrectly. The system s date and time setting are used to populate the DICOM header information on images. This could result is an inaccurate date recorded on the images.
-
- Terminated Z-0693-2019
Possible incorrect software version loaded.
-
- Terminated Z-3028-2017
Possible incorrect wiring of the Uninterruptable Power Supply, which could contribute to a potential shock hazard.
-
- Terminated Z-2266-2013
GE Healthcare has recently become aware of a potential safety issue due to non-unique series identifiers embedded in image headers of certain GE MR Products. For ClariView and Screen Save from GVTK (Gems Visualization ToolKit) Viewer, the MR Product can generate image headers with non-unique series identifiers, which when exported to some PACS systems, results in the series being associated wit
Specification
- Trade Name
- Na
- Generic Name
- Full-body MRI system, superconducting magnet
- Model / Reference
- G3341AA
- Primary DI / UDI-DI
- 00840682115964
- 510(k) Number
- FDA Product Code
- LNH, MOS, LNI
- GMDN Term
- Full-body MRI system, superconducting magnet
- Device Class
- Class II
- Regulation Number
- 892.1000
- Commercial Distribution
- In Commercial Distribution
Other devices from GE Medical Systems, L.L.C.
Properties & raw data
Show properties, registry sources and raw records
Other properties (11)
- Sterile: No
- Single Use: No
- Implantable: No
- Contains Latex: No
- Over-the-Counter: No
- Convenience Kit: No
- Combination Product: No
- HCT/P: No
- Premarket Exempt: No
- Lot / Batch Number: No
- Expiration Date: No
Sources (1)
- FDA UDI 780428a7-6459-4e42-a606-81bafb84c2f6