Na MIA16-8NS
Listed
Valid until
Not applicable
FDA recalls
9 ongoingDocuments
None
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | GUDID listing | Number00845384005155 | Date9 Sep 2016 | StatusListed |
Recalls
-
- Ongoing Z-0922-2023
- Ongoing Z-0917-2023
- Ongoing Z-0921-2023
- Ongoing Z-0923-2023
- Ongoing Z-0918-2023
- Ongoing Z-0920-2023
- Ongoing Z-0924-2023
- Ongoing Z-0919-2023
- Ongoing Z-0916-2023
Ball bearings in the CRANI-A (ANSPACH Power Tools Adult Craniotome Attachment) came out of the attachment, possibly during removal of the attachment, intra-operatively. Failure to follow the recommended inspection intervals provided in the IFU may result in serious patient injury
-
- Terminated Z-2013-2017
- Terminated Z-1951-2017
- Terminated Z-2020-2017
- Terminated Z-2028-2017
- Terminated Z-2000-2017
- Terminated Z-2025-2017
- Terminated Z-2022-2017
- Terminated Z-2027-2017
- Terminated Z-2012-2017
- Terminated Z-1932-2017
- Terminated Z-2014-2017
- Terminated Z-2030-2017
- Terminated Z-1986-2017
- Terminated Z-2002-2017
- Terminated Z-2024-2017
- Terminated Z-1933-2017
- Terminated Z-1987-2017
- Terminated Z-1953-2017
- Terminated Z-1990-2017
- Terminated Z-2023-2017
- Terminated Z-2029-2017
- Terminated Z-1930-2017
- Terminated Z-1950-2017
- Terminated Z-1985-2017
- Terminated Z-1947-2017
- Terminated Z-1954-2017
- Terminated Z-1988-2017
- Terminated Z-1994-2017
- Terminated Z-1946-2017
- Terminated Z-1931-2017
- Terminated Z-1958-2017
- Terminated Z-1991-2017
- Terminated Z-1999-2017
Supplied Directions for Use (DFU) did not contain a recommended service interval.
-
- Terminated Z-1574-2014
Anspach product code L-3SD contained product code S-3SD which is shorter in length than the labeled product.
-
- Terminated Z-0779-2014
Product did not reach expected sterility assurance level
-
- Terminated Z-1905-2014
- Terminated Z-1896-2014
- Terminated Z-1911-2014
- Terminated Z-1895-2014
- Terminated Z-1910-2014
- Terminated Z-1899-2014
- Terminated Z-1912-2014
- Terminated Z-1909-2014
- Terminated Z-1908-2014
- Terminated Z-1917-2014
- Terminated Z-1904-2014
- Terminated Z-1913-2014
The Anspach Effort, Inc. in Palm Beach Gardens, FL is recalling their Anspach Custom Devices. The units were not manufactured consistently with Quality System requirements.
-
- Terminated Z-0074-2013
- Terminated Z-0075-2013
- Terminated Z-0076-2013
The Anspach Effort, Inc. in Palm Beach Gardens, FL is recalling the foot control/foot pedal due to the Foot Control Cable lacking the shield connection from the cable to the connector body.
-
- Terminated Z-3191-2011
- Terminated Z-3189-2011
- Terminated Z-3190-2011
The Anspach Effort, Inc., Palm Beach Gardens, FL, is recalling their Anspach Single Use, Sterile Bone Cutting Burrs, catalog numbers, IMRI-S-8NS-M, M-6P, QD8-4OVD, S-5B-4, and SIL-7-9, lot #'s C323023789, C403026551, C473028567, C403026563, D153034013, D173034385, and C363024789, because the bar code which defines the expiration date may be incorrect. This could cause the bar code reader system to
-
- Terminated Z-0198-2014
Anspach Effort, Inc. Palm Beach Gardens, FL initiated a correction of its medical Device labeling for its Pneumatic Systems Operating Manual, Cutting Burr Chart and Product Catalog due to a recent review of their regulations and compliance requirements that determined the claim that the device is capable of dissection of biometals, bioplastics and cement has been verified, but is is not part of th
-
- Terminated Z-0830-2014
- Terminated Z-0829-2014
Cutter was longer than was stated on the label
-
- Terminated Z-3202-2011
- Terminated Z-3203-2011
The Anspach Effort, Palm Beach Gardens, FL, is recalling their Single Use, Sterile Bone Cutting Burrs, Catalog #, S-1504TD, Lot #s D363039140. D323038110, D293037349 and Catalog # TURQ-434-4, Lot #s D343038845, D323038131, due to the burr being 2.0 mm longer than stated on the label. As a result, a hole drilled using this burr would be 2 mm deeper than the screw.
-
- Terminated Z-0083-2010
The Anspach Effort, Inc. is recalling the Foot Control/ Foot Pedal for SC2000 and SC2100 Powered, simple cranial drills, burrs, trephines, and accessories. The Foot Control Cables lacks the shield connection from the cable to the connector body. The lack of shielding may result in radiated emissions which could possibly cause interference on other electrical equipment adjacent to the foot pedal
-
- Terminated Z-1008-2007
The white film portion of the sterile pouch may become brittle prior to the expiry date compromising the sterility of the product.
-
- Terminated Z-1243-03
Faulty component causing a malfunction which would unexpectedly shutdown the drill system and user may not be able to restart.
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — GUDID listing Listed
Specification
- Generic Name
- Cranial bur, single-use
- Model / Ref / Catalog
- MIA16-8NS
- Primary DI / UDI-DI
- 00845384005155
- 510(k) Number
- FDA Product Code
- HBC, ERL
- GMDN Term
- Cranial bur, single-use
- Device Class
- Class II
- Regulation Number
- 882.4360, 874.4250
- GUDID distribution status
- Not in Commercial Distribution
Other devices from The Anspach Effort, Inc.
Properties
Other properties (9)
- Implantable: No
- Contains Latex: No
- Over-the-Counter: No
- Convenience Kit: No
- Combination Product: No
- HCT/P: No
- Premarket Exempt: No
- Serial Number: No
- Manufacturing Date: No