Natural-Knee II System
510(k) K936159
Active
Valid until
No expiry
FDA recalls
None ongoingDocuments
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SN | NumberK936159 | Date22 May 1995 | StatusActive |
Recalls
-
- Terminated Z-1033-2018
The LDPE bag packaging for various highly polished hip and knee implants may adhere to the highly polished surface, leaving residue or material from the LDPE bag on the implant after it is removed from the bag.
-
- Terminated Z-0839-2016
LDPE bag containing the implant adheres to the highly polished implant surface.
-
- Terminated Z-1262-2014
- Terminated Z-1249-2014
A complaint was received for a discolored All Poly Tibia. Root cause analysis found that the discoloration was the result of free radical oxidation. The oxidation process was advanced prematurely as a result of the machined component being stored in a transparent container exposed to artificial light for six years prior to inert packaging and sterilization of the device.
-
- Terminated Z-0078-2014
- Terminated Z-0075-2014
The low density polyethylene (LDPE) bag used to package implants adheres to the highly polished surface of the devices.
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA Product Code
- JWH
- Device Class
- Class II
- Regulation
- 21 CFR Part 888.3560
- Advisory Committee
- Orthopedic
- FDA Decision
- SN