Skip to content

Nebulizer Heater

by Travenol Laboratories, S.A. 510(k)

Active United States

510(k) K780456

Regulatory status

as of 29 Oct 2026

RegionAuthorizationNumberDateStatus
United StatesFDA 510(k)SESE Substantially Equivalent · Traditional NumberK780456 Date12 Apr 1978ClearedStatusActive

Coverage

1 jurisdiction

United States of America
  • Active
  • Expiring
  • Withdrawn
  • Listed
  • Not covered

United States — FDA 510(k) Active

Recalls

1 recall · FDA

    • Terminated Z-3016-2011 FDA Unclassified (pre-2012) initiated 13 Jul 2011

    The recall was initiated because CareFusion's Investigation has confirmed the Nebulizer Heater can, over time, develop a failure mode that results in an internal electrical short. The short can result in smoke and sparks being emitted from the device. In the presence of oxygen enriched environments and a fuel source, this failure represents a potential risk of flame. CareFusion is voluntarily rec

Specification

510(k)

510(k) Number
FDA Product Code
CAF
Device Class
Class II
Regulation
21 CFR Part 868.5630
Advisory Committee
Anesthesiology
FDA Decision
SESE Substantially Equivalent

Other devices from Travenol Laboratories, S.A.

See all devices from Travenol Laboratories, S.A. →