Nemio Diagnostic Ultrasound System, Model Ssa-550A

by Toshiba America Medical Systems, In.C

510(k) K043078

Active United States

510(k)

Valid until

No expiry

510(k) clearance is not time-limited

FDA recalls

None ongoing

1 closed

Documents

Trust

Unverified

Registry data, not yet checked by deviCERTS. Source FDA 510(k), synced 22 Oct 2026. What this means

Regulatory status

as of registry sync 22 Oct 2026

RegionAuthorizationNumberDateStatus
United StatesFDA 510(k)SESE Substantially Equivalent · Traditional NumberK043078 Date17 Nov 2004ClearedStatusActive

Recalls

1 recall · FDA

    • Terminated Z-0253-2008 FDA Unclassified (pre-2012) initiated 29 Aug 2007

    Multiple problems including : (1) The surface temperature of the transducer tip can rise and exceed the limit set by the international standard. (2) Incorrect display of the TI indices on the display monitor. The displayed value is greater or smaller than the true TI value, and (3) Manuals contain incorrect values for MI/TI values.

Coverage

1 jurisdiction

United States of America
  • Active
  • Expiring
  • Withdrawn
  • Listed
  • Not covered

United States — FDA 510(k) Active

Specification

510(k)

510(k) Number
FDA Product Code
IYO
Device Class
Class II
Regulation
21 CFR Part 892.1560
Advisory Committee
Radiology
FDA Decision
SESE Substantially Equivalent