Nemio Xg, Saa-580A
by Toshiba America Medical Systems, In.C
510(k) K061605
Active
Valid until
No expiry
FDA recalls
None ongoingDocuments
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK061605 | Date22 Jun 2006 | StatusActive |
Recalls
-
- Terminated Z-0254-2008
Multiple problems including : (1) The surface temperature of the transducer tip can rise and exceed the limit set by the international standard. (2) Incorrect display of the TI indices on the display monitor. The displayed value is greater or smaller than the true TI value, and (3) Manuals contain incorrect values for MI/TI values.
-
- Terminated Z-0771-2007
Transducer Surface Temperature : It was found during testing that when using the system in the 2DCDI + PW mode, with the Scale Graph function active, that the transducer surface temperature could exceed the 27 degrees Celsius limit as specified in the IEC safety standard.
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA Product Code
- IYN
- Device Class
- Class II
- Regulation
- 21 CFR Part 892.1550
- Advisory Committee
- Radiology
- FDA Decision
- SESE Substantially Equivalent