Neurosign 100 Unit
by The Magstim Company Limited
Ref 9883-01
Neurosign 100 Unit (9883-01), a nerve-locating system (EU MDD Class IIa; FDA Class II; FDA product code ETN, Stimulator, Nerve) made by The Magstim Company Limited, was registered in EUDAMED on 25 Feb 2022 (Basic UDI-DI B-05060229530943), cleared under FDA 510(k) K923056 on 24 Nov 1992 and listed in FDA GUDID on 23 Sep 2016 (UDI-DI 05060229530943); the registry lists it as withdrawn (EUDAMED) as of 25 Oct 2026. FDA recalls on record: 1.
Active
Valid until
No expiry
FDA recalls
None ongoingTrust
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| European Union | CE/MDD EUDAMED | NumberB-05060229530943 05060229530943 | Date25 Feb 2022 | StatusWithdrawn |
| United States | GUDID listing GUDID | Number05060229530943 | Date23 Sep 2016 | StatusListed |
| United States | FDA 510(k)SESE GUDID | NumberK923056 | Date24 Nov 1992 | StatusActive |
Official records
- EUDAMED public site B-05060229530943
- AccessGUDID record 05060229530943
- FDA 510(k) database K923056
From the FDA record
- Device description
- Intraoperative nerve monitor
- MRI safety
- Labeling does not contain MRI Safety Information
- Natural rubber latex
- No latex statement on the label
- Sterility
- Supplied non-sterile
Across registries
- EUDAMED Basic UDI-DI B-05060229530943 registered 25 Feb 2022 Withdrawn
- FDA 510(k) K923056 decided 24 Nov 1992 Active
- FDA GUDID UDI-DI 05060229530943 listed 23 Sep 2016 Listed
Recalls
-
- Terminated Z-0869-2021
A small number of Neurosign 100 Intraoperative Nerve Monitors may have been dispatched without being configured appropriately for the supply mains within the target Country. The specific deficiency includes: 1. Onboard voltage selector set to 240V (European Standard), versus 120V requirement for the United States and Canada. 2. Fuses installed in the Power Entry Module (PEM) 2xT315mAL (240V Standard), versus the 2xT630mAL requirement for a 120V mains supply.
Timeline
- Cleared FDA 510(k) K923056
- Listed FDA GUDID 05060229530943
- Recall initiated FDA recalls Z-0869-2021
- Registered EUDAMED B-05060229530943
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
- European Union
- United States
Specification
- Trade name
- Neurosign 100 Unit EUDAMEDNeurosign 100 GUDID
- Generic name
- Nerve-locating system, line-powered
- Model / reference
- 9883-01
- Primary DI / UDI-DI
- 05060229530943
- Basic UDI-DI
- B-05060229530943
- 510(k) Number
- FDA product code
- ETN
- EMDN code
- Z121002 Nervous system monitoring systems
- GMDN term
- Nerve-locating system, line-powered
- Device class
- Class IIa EUDAMEDClass II GUDID
- Regulation number
- 874.1820
- System or procedure pack
- System which is a device in itself
- EU registration type
- Legacy device (under a directive)
- EU market of registration
- Netherlands
- GUDID distribution status
- In Commercial Distribution
Truth per registry
| Field | EUDAMED | GUDID |
|---|---|---|
| Device class | Class IIa | Class II |
Manufacturer details
- EUDAMED SRN
- GB-MF-000003956
- Authorised representative
- Technomed Europe NL-AR-000002048
Similar devices in this category
Other devices filed under NERVOUS SYSTEM MONITORING SYSTEMS.
- Friendship Medical 3019 by Xi'an Friendship Medical Electronics Co., Ltd.
- Cadwell 0021 by Xi'an Friendship Medical Electronics Co., Ltd.
- Stimulationskabel für TWISTER® TC, 12 Kontakte 20-0202 by Dr. Langer Medical GmbH
- ISIS 504281 by inomed Medizintechnik GmbH
- Cascade IOMAX Intraoperative Monitor by Cadwell Industries, Inc.
- Neurostimulator TWISTER® TC 20-0200 by Dr. Langer Medical GmbH
- EpiMonitor by Empatica EpiMonitor-AND by Empatica S.r.L.
- EpiMonitor by Empatica EpiMonitor-IOS by Empatica S.r.L.
- ISIS 504261 by inomed Medizintechnik GmbH
- Cadwell Guardian 199332-000 by Cadwell Industries, Inc.