# Neurosign Model 100

> Active: United States. Medical device record on deviCERTS, compiled from public registries.

- Canonical page: <https://devicerts.com/devices/neurosign-model-100-etn>
- Manufacturer: [The Magstim Company Limited](https://devicerts.com/manufacturers/the-magstim-company-limited) (United Kingdom)
- Name in the registry: NEUROSIGN MODEL 100
- Device class: Class II (US FDA)
- FDA product code: ETN

## Registrations and authorizations

- **510(k)** · FDA 510(k) · United States · K923056 · decision 1992-11-24 · decision SESE · status: Active

## Recalls

- Z-0869-2021 · Class II · terminated · initiated 2020-12-08: A small number of Neurosign 100 Intraoperative Nerve Monitors may have been dispatched without being configured appropriately for the supply mains within the target Country. The specific deficiency includes: 1. Onboard voltage selector set to 240V (European Standard), versus 120V requirement for the United States and Canada. 2. Fuses installed in the Power Entry Module (PEM) 2xT315mAL (240V Standard), versus the 2xT630mAL requirement for a 120V mains supply.

## Identifiers

- 510(k) Number: K923056 [FDA 510(k)]
- FDA product code: ETN [FDA 510(k)]

## Classification

- Device class: Class II [FDA 510(k)]
- Regulation: 21 CFR Part 874.1820 [FDA 510(k)]
- Advisory committee: Ear, Nose, Throat [FDA 510(k)]
- FDA decision: SESE [FDA 510(k)]

## Source and citation

- Cite as: Neurosign Model 100, The Magstim Company Limited. deviCERTS medical device catalogue. https://devicerts.com/devices/neurosign-model-100-etn (data as of 2026-11-27).
- Data sources: [FDA 510(k)](https://open.fda.gov/apis/device/510k/), [FDA recalls](https://open.fda.gov/apis/device/recall/)
- Registry records last synced: 2026-11-13
- Statuses computed as of: 2026-11-27
- JSON: <https://devicerts.com/devices/neurosign-model-100-etn.json>
- Licence: [CC BY 4.0](https://creativecommons.org/licenses/by/4.0/). The deviCERTS layout and text are licensed under CC BY 4.0. Registry data remain under their source terms: U.S. FDA data (openFDA, AccessGUDID) are U.S. public domain; EUDAMED data are subject to the European Commission reuse terms.
- Attribution: DevicErts, https://devicerts.com/devices/neurosign-model-100-etn
- Compiled by deviCERTS (devicerts.com) from public medical device registries (EUDAMED, European Commission; U.S. FDA via openFDA; AccessGUDID, U.S. National Library of Medicine). deviCERTS is not an official register and gives no regulatory advice: the record in the registry itself prevails.