Skip to content

Newdeal Ankle Nail

by New Deal, S.A.

510(k) K050882

Active United States

510(k)

Valid until

No expiry

510(k) clearance is not time-limited

FDA recalls

None ongoing

5 closed

Documents

Trust

Unverified

Registry data, not yet checked by deviCERTS. Source FDA 510(k), synced 22 Oct 2026. What this means

Regulatory status

as of registry sync 22 Oct 2026

RegionAuthorizationNumberDateStatus
United StatesFDA 510(k)SESE Substantially Equivalent · Traditional NumberK050882 Date27 May 2005ClearedStatusActive

Recalls

5 recalls · FDA

    • Terminated Z-1539-2017 FDA Class II initiated 3 Mar 2017

    Voluntary recall/Removal of Panta and Panta XI Nails because there may be the potential that the internal and external package seal does not provide the level of sterility intended.

    • Terminated Z-1736-2013 FDA Class II initiated 9 May 2013

    As a result of the investigation into an adverse trend of complaints for reports of misalignment of the calcaneal or tibial screws during use of the Panta Nail Support Device, Integra has decided to issue this voluntary medical device correction. The misalignment can result in the inability to properly insert the screws into the Panta Nail and create the potential need for revision surgery.

    • Terminated Z-1428-2012 FDA Unclassified (pre-2012) initiated 19 May 2011

    The firm received multiple complaints from their user customers (surgeons) who reported encountering difficulty inserting a calcaneal or tibial screw through the holes of the PANTA nail.

    • Terminated Z-0308-2011 FDA Unclassified (pre-2012) initiated 1 Mar 2010

    Inability to properly align the calcaneal screw placement on a consistent basis, due to a manufacturing defect of of the targeting frame (Panta Support Device).

    • Terminated Z-1877-2008 FDA Unclassified (pre-2012) initiated 24 Mar 2008

    Certain PANTA Nails from one lot were incorrectly manufactured. The internal thread for certain PANTA Nails (Part Number 500050ND; Lot Number: E7XX) may present an insufficient depth and may not allow the engagement of the threaded part of the compression device. The risks presented include: an increase in surgery time, or the implantation of a nail less ideally suited for the patient's anatomy

Coverage

1 jurisdiction

United States of America
  • Active
  • Expiring
  • Withdrawn
  • Listed
  • Not covered

United States — FDA 510(k) Active

Specification

510(k)

510(k) Number
FDA Product Code
HSB
Device Class
Class II
Regulation
21 CFR Part 888.3020
Advisory Committee
Orthopedic
FDA Decision
SESE Substantially Equivalent

Other devices from New Deal, S.A.

See all devices from New Deal, S.A. →